TABLE 7.

Image Acquisition Parameters (3,5)

Dynamic flowStaticSPECT
AnteriorAnterior for blood pool images; additional static images include posterior and lateralsOnly if using 99mTc-HMPAO or 99mTc-ECD and only as needed
1–3 s/frame for 1–2 min500,000–1,000,000 counts per imageTime per stop and counts acquired dependent on radiopharmaceutical dose administered (e.g., 64 projections, 20–40 s/stop [200,000 counts]); use of institutional guideline for other brain SPECT studies
15%–20% window around 140-keV photopeak15%–20% window around 140-keV photopeak15%–20% window around 140-keV photopeak
128 × 128 matrix128 × 128 matrixAt least 128 × 128 matrix
LEAP, LEHR collimatorsLEAP, LEHR collimatorsLEAP, LEHR, or UHR collimators; fanbeam or other focused collimator is needed for increased resolution and higher counting rates
Zooming or magnification optionalZooming or magnification optionalZoom set to produce pixel size of 3.5 mm or less
Circular or noncircular orbit with smallest radius possible
Continuous or step-and-shoot
  • LEAP = low-energy all-purpose; LEHR = low-energy high-resolution; UHR = ultrahigh resolution.