TABLE 5.

Results of Cardiac Function Analysis of Myocardial Wall Variation

Parameter20:2020:1020:05
ED (mL)ES (mL)LVEF (%)ED (mL)ES (mL)LVEF (%)ED (mL)ES (mL)LVEF (%)
Theoretical phantom values (mL)556.9216.761.1479.9216.754.8441.5216.750.9
CT421.9186.755.7397.4192.151.7385.4193.549.8
STPG SPECT
 STPGAC420.0187.055.5396.0191.051.8385.0194.049.6
 UnTagAC42318656.039118552.738619449.7
 75% GAC42218655.940019351.838619449.7
 40% GAC4251875640619651.738819450.0
NPG SPECT
 UnTagAC423.0186.056.0391.0185.052.7386.0194.049.7
 75% GAC422.0186.055.9400.0193.051.8386.0194.049.7
 40% GAC425.0187.056.0406.0196.051.7388.0194.050.0
 Non-AC418.0191.054.3382.0185.051.6388.0201.048.2
  • Non-AC = non—attenuation-corrected.