TABLE 1.

Cardiac Function Analysis of Heart Rate Variability (360°)

ParameterTheoretic phantom values (mL)CTSTPG SPECTNPG SPECT
STPGACUnTagAC75% GAC40% GACUnTagAC75% GAC40% GACNon-AC
40 bpm
 ED (mL)140.4128.1128.0127.0127.0129.0130.0129.0131.0130.0
 ES (mL)119.9112.0112.0107.0108.0109.0112.0111.0113.0112.0
 LVEF (%)14.612.612.515.715.015.513.814.013.713.8
50 bpm
 ED (mL)140.4129.5130.0128.0129.0130.0131.0131.0133.0131.0
 ES (mL)119.9114.6115.0111.0112.0112.0114.0114.0117.0115.0
 LVEF (%)14.611.511.513.313.213.813.013.012.012.2
60 bpm
 ED (mL)140.4128.8128.0124.0125.0127.0127.0127.0129.0127.0
 ES (mL)119.9111.6111.0105.0106.0108.0108.0110.0111.0111.0
 LVEF (%)14.613.413.315.315.215.015.013.414.012.6
70 bpm
 ED (mL)140.4132.6133.0128.0129.0130.0132.0133.0133.0131.0
 ES (mL)119.9115.7116.0108.0109.0111.0112.0113.0114.0113.0
 LVEF (%)14.612.712.815.615.514.615.215.014.313.7
80 bpm
 ED (mL)140.4133.2133.0129.0130.0131.0131.0133.0135.0133.0
 ES (mL)119.9116.8116.0111.0112.0113.0113.0115.0117.0116.0
 LVEF (%)14.612.312.814.013.813.713.713.513.312.8
90 bpm
 ED (mL)140.4131.9132.0127.0129.0130.0132.0132.0134.0130.0
 ES (mL)119.9114.9115.0108.0110.0110.0113.0113.0115.0111.0
 LVEF (%)14.612.912.915.014.715.414.414.414.214.6
100 bpm
 ED (mL)140.4131.4131.0128.0129.0129.0131.0131.0134.0130.0
 ES (mL)119.9115.5115.0110.0111.0112.0114.0113.0117.0113.0
 LVEF (%)14.612.112.214.114.013.213.013.712.713.1
  • Non-AC = non—attenuation-corrected.