Prescan | | |
Preparation | No specific preparation; no fasting required | |
Scan | Rest scan | Standard |
Dose | 370–740 MBq (10–20 mCi) 99mTc-HMDP intravenously | Standard |
Time from injection to acquisition | 2 or 3 h | Standard |
General imaging parameters | | |
Field of view | Heart or chest | Standard |
| Whole-body | Optional |
Image type | Planar | Standard |
| SPECT if planar is positive | Required |
Position | Supine | Standard |
| Upright | Optional |
Energy window | 140 keV, 15%–20% | Standard |
Collimators | Low-energy, high-resolution | Standard |
Matrix | Planar: 256 × 256 (minimum, 64 × 64) | Standard |
| SPECT: 128 × 128 (minimum, 64 × 64) | Standard |
Pixel size | 3.5–6.5 mm | Standard |
Planar parameters | | |
Views | Chest: anterior and lateral | Standard |
| Whole-body: from head to toe | Optional |
Detector configuration | 90° | Recommended |
Image duration (count-based) | Planar: 750,000 | Standard |
| Whole-body: 20 cm/min | Optional |
Magnification | 1.46 | Standard |
SPECT parameters | | |
Angular range | 180° | Standard |
| 360° | Optional |
Detector configuration | 90° | Standard |
| 180° | Optional |
Electrocardiographic gating | Off; nongated imaging | Standard |
Number of views/detector | 40/32 | Standard |
Time per stop | 20 s/25 s | Standard |
Magnification | 1.46 (180° angular range) or 1.0 (360° angular range) | Recommended |