Quality Control Test Results for Eluted Solutions and Final Products
Test | Accepted limit | Average result |
---|---|---|
99Mo breakthrough | <0.1% | 0.02% |
Al+3 breakthrough | Eluted spot less intensely colored than standard solution | All passed |
99mTc pH | 4.0–8.0 | 5.0 |
99mTc-DMSA pH | 2.5–3.5 | 3.1 |
Labeling efficiency | ≥95% | 98.2% |