TABLE 2.

Quality Control Test Results for Eluted Solutions and Final Products

TestAccepted limitAverage result
99Mo breakthrough<0.1%0.02%
Al+3 breakthroughEluted spot less intensely colored than standard solutionAll passed
99mTc pH4.0–8.05.0
99mTc-DMSA pH2.5–3.53.1