TABLE 1.

Relationship Between Bed Motion Speed, Radioactivity Concentration, and SUV in 68Ge Cylinder Phantom

Bed speed (mm/s)Activity (kBq/mL)SUVmeanSUV SDSUVmax
StaticOncoFreeze AIStaticOncoFreeze AIStaticOncoFreeze AIStaticOncoFreeze AI
0.69.059.031.041.040.040.041.221.28
0.99.069.021.041.040.050.061.251.33
1.29.059.021.041.040.050.061.311.39
1.59.079.041.041.040.060.071.341.38
1.89.069.021.041.040.070.081.441.50
2.19.089.041.051.040.070.091.441.50
2.49.069.001.041.040.080.101.421.55
2.79.079.011.041.040.080.101.471.52
3.09.078.991.041.040.090.101.431.59