Relationship Between Bed Motion Speed, Radioactivity Concentration, and SUV in 68Ge Cylinder Phantom
Bed speed (mm/s) | Activity (kBq/mL) | SUVmean | SUV SD | SUVmax | ||||
---|---|---|---|---|---|---|---|---|
Static | OncoFreeze AI | Static | OncoFreeze AI | Static | OncoFreeze AI | Static | OncoFreeze AI | |
0.6 | 9.05 | 9.03 | 1.04 | 1.04 | 0.04 | 0.04 | 1.22 | 1.28 |
0.9 | 9.06 | 9.02 | 1.04 | 1.04 | 0.05 | 0.06 | 1.25 | 1.33 |
1.2 | 9.05 | 9.02 | 1.04 | 1.04 | 0.05 | 0.06 | 1.31 | 1.39 |
1.5 | 9.07 | 9.04 | 1.04 | 1.04 | 0.06 | 0.07 | 1.34 | 1.38 |
1.8 | 9.06 | 9.02 | 1.04 | 1.04 | 0.07 | 0.08 | 1.44 | 1.50 |
2.1 | 9.08 | 9.04 | 1.05 | 1.04 | 0.07 | 0.09 | 1.44 | 1.50 |
2.4 | 9.06 | 9.00 | 1.04 | 1.04 | 0.08 | 0.10 | 1.42 | 1.55 |
2.7 | 9.07 | 9.01 | 1.04 | 1.04 | 0.08 | 0.10 | 1.47 | 1.52 |
3.0 | 9.07 | 8.99 | 1.04 | 1.04 | 0.09 | 0.10 | 1.43 | 1.59 |