TABLE 2.

Comparison of A+B/G and D+F/G Ratios Between Carbon and XPS Headrests in Patients with Cerebrovascular Disease (n = 43)

A+B/GD+F/G
ParameterCarbonXPSPCarbonXPSP
FBP Chang AC0.96 ± 0.050.91 ± 0.05<0.001*0.87 ± 0.040.87 ± 0.040.343
OSEM Chang AC0.96 ± 0.050.92 ± 0.06<0.001*0.88 ± 0.040.88 ± 0.040.938
OSEM CTAC0.89 ± 0.040.89 ± 0.050.8930.86 ± 0.030.86 ± 0.030.099
  • * Statistically significant.