TABLE 1.

Typical Specifications for Quality Control Tests

TestSpecification
Radiochemical purity (HPLC)>95%
Radiochemical identity (HPLC)tR ± 5% reference standard
Appearance (visual inspection)Clear, colorless, particulate-free
pH4.0–7.0
Endotoxin content (USP <85>)<175 EU/injected dose
Filter integrity (bubble point)According to filter manufacturer
SterilitySterile after 14 d