TABLE 1.

Thyroid Uptake Using 131I Capsules for 3 Patients

Dose calibratorProbe24-h uptake
Patient no.Activity (MBq)TimeCount 1 (cps)Count 2 (cps)Room background counts (cps)Net capsule counts (cps)TimeRoom background timeProbe efficiency (cps/MBq)Original (%)Adjusted (%)
10.4209:1621321012118:268:23501139.6%71%
20.43510:3521421612148:298:2349034.8%18%
30.3889:08208206120613:248:2353447.0%24%
  • Original 24-h thyroid uptake results are shown along with their corresponding adjusted values after compensating for deviations in probe efficiencies from previous patients (Table 2). Time is expressed as time of day (hr:min).