TABLE 1.

Fit Parameters for Time to Elute Certain Activity Levels as Function of Generator Age, Following Functional Form Formula

Fit (mCi)y intercept (y0) (s)Amplitude (A) (s)Growth constant (t1) (d−1)Band range (s)
206.052.3137.54±1.05
306.255.2132.60±1.70
409.255.7017.33±5
452.9313.1220±8
Modified 4516.223.1510.96±2.5
  • Additional fit was performed for largest eluted activity (1,665 MBq) to demonstrate that data could be more precisely fitted without 2 days’ worth of generator results.