TY - JOUR T1 - Extrinsic Versus Intrinsic Uniformity Correction for γ-Cameras JF - Journal of Nuclear Medicine Technology JO - J. Nucl. Med. Technol. SP - 208 LP - 212 DO - 10.2967/jnmt.110.084814 VL - 39 IS - 3 AU - Randy Bolstad AU - Jody Brown AU - Vesper Grantham Y1 - 2011/09/01 UR - http://tech.snmjournals.org/content/39/3/208.abstract N2 - SPECT produces nuclear medicine images using a 3-dimensional diagnostic tool that eliminates the superimposition of adjacent structures, thus providing improved disease localization. Another method of uniformity correction—to use the complete capabilities of this tool—is discussed and evaluated in this article. The conventional method of intrinsic uniformity correction accounts only for nonuniformities within the γ-camera, excluding the collimator. If SPECT image quality is related to overall camera performance, then using an extrinsic uniformity correction method rather than an intrinsic method will improve image quality. Methods: SPECT uniformity images were obtained using a SPECT phantom with application of intrinsic and extrinsic uniformity correction tables with 2 different γ-imaging systems. The image results were qualitatively assessed. Results: Even with acceptable nonuniformity analyses, significant ring artifacts within the intrinsic uniformity–corrected images are observed, whereas the artifacts are considerably less significant with the extrinsic uniformity correction and disappear completely in some of these images. Conclusion: Extrinsic uniformity correction may significantly improve the overall image quality by taking into account nonuniformities that arise from the collimator. This method will result in fewer image artifacts and improved image quality, thereby improving patient care. ER -