PT - JOURNAL ARTICLE AU - Rasouli, Mahmoud AU - Takavar, Abbas AU - Ay, Mohammad Reza AU - Saber, Sasan AU - Loudos, George TI - Effects of Crystal Pixel Size and Collimator Geometry on the Performance of a Pixelated Crystal γ-Camera Using Monte Carlo Simulation AID - 10.2967/jnmt.109.073072 DP - 2010 Dec 01 TA - Journal of Nuclear Medicine Technology PG - 199--204 VI - 38 IP - 4 4099 - http://tech.snmjournals.org/content/38/4/199.short 4100 - http://tech.snmjournals.org/content/38/4/199.full SO - J. Nucl. Med. Technol.2010 Dec 01; 38 AB - Dedicated γ-cameras based on pixelated scintillators have long been used for breast tumor imaging. Intercrystal scattering (ICS) increases the background counting rate and degrades the image quality when small crystal pixels are used. Because of the small size of applied collimators, scattered radiation and septal penetration are high, and therefore collimator characteristics must be carefully considered. In our study, we investigated the influence of ICS events on position-detection accuracy (PDA) for pixelated crystals and the effects of different geometries of hexagonal-hole collimators on the performance of these cameras, using Monte Carlo simulation to optimize camera design. The arrays of thallium-doped cesium iodide detectors with different pixel dimensions that had been exposed to 140-keV photons of isotropic point source, 50 mm from the collimator surface, were simulated. Hexagonal-hole collimators were 10.5, 15, and 21 mm long. The septal thickness varied from 0.1 to 0.5 mm, with 3 different hole diameters. The results confirmed that by increasing the crystal pixel size, ICS was decreased and change of detection efficiency was negligible, but PDA, contrast-to-noise ratio, and spatial resolution (full width at half maximum) were increased. Our experiences confirmed that 2 × 2 mm was an optimum crystal pixel size, especially for a lower ICS fraction and an appropriate full width at half maximum. Because collimators are the limiting factor for spatial resolution and sensitivity, careful collimator design is of great importance.