RT Journal Article SR Electronic T1 Nonuniformity Intrinsic Flood Artifact Caused by Point-Source Syringe Needle JF Journal of Nuclear Medicine Technology JO J. Nucl. Med. Technol. FD Society of Nuclear Medicine SP 41 OP 43 VO 25 IS 1 A1 Hackett, Michael T. A1 Magoun, Sylvia L. A1 Thompson, Rosemary B. YR 1997 UL http://tech.snmjournals.org/content/25/1/41.abstract AB The purpose of this study was to investigate the cause of a nonuniformity artifact seen on an intrinsic flood. Methods: Intrinsic floods were acquired on three different cameras using a point source in a 1-cc syringe, with and without a needle attached. Floods were obtained with the syringe placed perpendicular and parallel to the camera face. Results: A nonuniformity artifact was evident if a syringe with a needle was placed perpendicular to the crystal face. Conclusion: If a syringe is used as a point source during the acquisition of an intrinsic quality control flood, the needle should not be placed perpendicular to the camera face.