PT - JOURNAL ARTICLE AU - Charest, Mathieu AU - Asselin, Chantal TI - Effective Dose in Nuclear Medicine Studies and SPECT/CT: Dosimetry Survey Across Quebec Province AID - 10.2967/jnmt.117.202879 DP - 2018 Jun 01 TA - Journal of Nuclear Medicine Technology PG - 107--113 VI - 46 IP - 2 4099 - http://tech.snmjournals.org/content/46/2/107.short 4100 - http://tech.snmjournals.org/content/46/2/107.full SO - J. Nucl. Med. Technol.2018 Jun 01; 46 AB - The aims of the current study were to draw a portrait of the delivered dose in selected nuclear medicine studies in Québec province and to assess the degree of change between an earlier survey performed in 2010 and a later survey performed in 2014. Methods: Each surveyed nuclear medicine department had to complete 2 forms: the first, about the administered activity in selected nuclear medicine studies, and the second, about the CT parameters used in SPECT/CT imaging, if available. The administered activities were converted into effective doses using the most recent conversion factors. Diagnostic reference levels were computed for each imaging procedure to obtain a benchmark for comparison. Results: The distributions of administered activity in various nuclear medicine studies, along with the corresponding distribution of the effective doses, were determined. Excluding 131I for thyroid studies, 67Ga-citrate for infectious workups, and combined stress and rest myocardial perfusion studies, the remainder of the 99mTc-based studies delivered average effective doses clustered below 10 mSv. Between the 2010 survey and the 2014 survey, there was a statistically significant decrease in delivered dose from 18.3 to 14.5 mSv. 67Ga-citrate studies for infectious workups also showed a significant decrease in delivered dose from 31.0 to 26.2 mSv. The standardized CT portion of SPECT/CT studies yielded a mean effective dose 14 times lower than the radiopharmaceutical portion of the study. Conclusion: Between 2010 and 2014, there was a significant decrease in the delivered effective dose in myocardial perfusion and 67Ga-citrate studies. The CT portions of the surveyed SPECT/CT studies contributed a relatively small fraction of the total delivered effective dose.