RT Journal Article SR Electronic T1 A Novel 141Ce-Based Flood Field Phantom: Assessment of Suitability for Daily Uniformity Testing in a Clinical Nuclear Medicine Department JF Journal of Nuclear Medicine Technology JO J. Nucl. Med. Technol. FD Society of Nuclear Medicine SP 225 OP 229 DO 10.2967/jnmt.116.189340 VO 45 IS 3 A1 Ashish Kumar Jha A1 Sneha Mithun A1 Manoj H. Chauhan A1 Nilendu Purandare A1 Sneha Shah A1 Archi Agrawal A1 Sanjay Kumar Saxena A1 Ashutosh Dash A1 Venkatesh Rangarajan YR 2017 UL http://tech.snmjournals.org/content/45/3/225.abstract AB Daily quality control testing of a γ-camera is of the utmost importance in assessing whether the camera is suitable for clinical use. The aim of our study was to assess the suitability of a fillable 141Ce-based flood field phantom developed in-house for daily quality control testing of γ-cameras. Methods: Daily uniformity testing was performed for 113 d using the fillable 141Ce phantom and a commercially available sheet-type 57Co phantom, and the results were compared. Results: The average integral uniformity obtained by the 141Ce and 57Co phantoms was 3.24% and 2.72%, respectively, for detector 1 and 3.31% and 2.78%, respectively, for detector 2. Conclusion: The 141Ce phantom we developed is a suitable alternative to the commercially available 57Co phantom.