PT - JOURNAL ARTICLE AU - Ashish Kumar Jha AU - Sneha Mithun AU - Manoj H. Chauhan AU - Nilendu Purandare AU - Sneha Shah AU - Archi Agrawal AU - Sanjay Kumar Saxena AU - Ashutosh Dash AU - Venkatesh Rangarajan TI - A Novel <sup>141</sup>Ce-Based Flood Field Phantom: Assessment of Suitability for Daily Uniformity Testing in a Clinical Nuclear Medicine Department AID - 10.2967/jnmt.116.189340 DP - 2017 Sep 01 TA - Journal of Nuclear Medicine Technology PG - 225--229 VI - 45 IP - 3 4099 - http://tech.snmjournals.org/content/45/3/225.short 4100 - http://tech.snmjournals.org/content/45/3/225.full SO - J. Nucl. Med. Technol.2017 Sep 01; 45 AB - Daily quality control testing of a γ-camera is of the utmost importance in assessing whether the camera is suitable for clinical use. The aim of our study was to assess the suitability of a fillable 141Ce-based flood field phantom developed in-house for daily quality control testing of γ-cameras. Methods: Daily uniformity testing was performed for 113 d using the fillable 141Ce phantom and a commercially available sheet-type 57Co phantom, and the results were compared. Results: The average integral uniformity obtained by the 141Ce and 57Co phantoms was 3.24% and 2.72%, respectively, for detector 1 and 3.31% and 2.78%, respectively, for detector 2. Conclusion: The 141Ce phantom we developed is a suitable alternative to the commercially available 57Co phantom.