RT Journal Article SR Electronic T1 Transmission Computed Tomography Data Acquisition with a SPECT System JF Journal of Nuclear Medicine Technology JO J. Nucl. Med. Technol. FD Society of Nuclear Medicine SP 53 OP 56 VO 15 IS 2 A1 Greer, Kim L. A1 Harris, C. Craig A1 Jaszczak, Ronald J. A1 Coleman, R. Edward A1 Hedlund, Laurence W. A1 Floyd, Carey E. A1 Manglos, Stephen H. YR 1987 UL http://tech.snmjournals.org/content/15/2/53.abstract AB Phantom and animal transmission computed tomography (TCT) scans were performed with a camera-based single photon emission computed tomography (SPECT) system to determine system linearity as a function of object density, which is important in the accurate determination of attenuation coefficients for SPECT attenuation compensation. Results from phantoms showed promise in providing a linear relationship (R = 0.997) in measuring density while maintaining good image resolution (FWHM = 10.5 mm). Animal images were essentially free of artifacts. Transmission computed tomography scans derived from a SPECT system appear to have the potential to provide data suitable for incorporation in an attenuation compensation algorithm at relatively low (calculated) radiation doses to the subjects.