PT - JOURNAL ARTICLE AU - Greer, Kim L. AU - Harris, C. Craig AU - Jaszczak, Ronald J. AU - Coleman, R. Edward AU - Hedlund, Laurence W. AU - Floyd, Carey E. AU - Manglos, Stephen H. TI - Transmission Computed Tomography Data Acquisition with a SPECT System DP - 1987 Jun 01 TA - Journal of Nuclear Medicine Technology PG - 53--56 VI - 15 IP - 2 4099 - http://tech.snmjournals.org/content/15/2/53.short 4100 - http://tech.snmjournals.org/content/15/2/53.full SO - J. Nucl. Med. Technol.1987 Jun 01; 15 AB - Phantom and animal transmission computed tomography (TCT) scans were performed with a camera-based single photon emission computed tomography (SPECT) system to determine system linearity as a function of object density, which is important in the accurate determination of attenuation coefficients for SPECT attenuation compensation. Results from phantoms showed promise in providing a linear relationship (R = 0.997) in measuring density while maintaining good image resolution (FWHM = 10.5 mm). Animal images were essentially free of artifacts. Transmission computed tomography scans derived from a SPECT system appear to have the potential to provide data suitable for incorporation in an attenuation compensation algorithm at relatively low (calculated) radiation doses to the subjects.