RT Journal Article SR Electronic T1 A Monte Carlo Study of the Dependence of Early Frame Sampling on Uncertainty and Bias in Pharmacokinetic Parameters from Dynamic PET JF Journal of Nuclear Medicine Technology JO J. Nucl. Med. Technol. FD Society of Nuclear Medicine SP 53 OP 60 DO 10.2967/jnmt.114.141754 VO 43 IS 1 A1 Ida Häggström A1 Jan Axelsson A1 Charles Ross Schmidtlein A1 Mikael Karlsson A1 Anders Garpebring A1 Lennart Johansson A1 Jens Sörensen A1 Anne Larsson YR 2015 UL http://tech.snmjournals.org/content/43/1/53.abstract AB Compartmental modeling of dynamic PET data enables quantification of tracer kinetics in vivo, through the calculated model parameters. In this study, we aimed to investigate the effect of early frame sampling and reconstruction method on pharmacokinetic parameters obtained from a 2-tissue model, in terms of bias and uncertainty (SD). Methods: The GATE Monte Carlo software was used to simulate 2 × 15 dynamic 3′-deoxy-3′-18F-fluorothymidine (18F-FLT) brain PET studies, typical in terms of noise level and kinetic parameters. The data were reconstructed by both 3-dimensional (3D) filtered backprojection with reprojection (3DRP) and 3D ordered-subset expectation maximization (OSEM) into 6 dynamic image sets with different early frame durations of 1, 2, 4, 6, 10, and 15 s. Bias and SD were evaluated for fitted parameter estimates, calculated from regions of interest. Results: The 2-tissue-model parameter estimates K1, k2, and fraction of arterial blood in tissue depended on early frame sampling, and a sampling of 6–15 s generally minimized bias and SD. The shortest sampling of 1 s yielded a 25% and 42% larger bias than the other schemes, for 3DRP and OSEM, respectively, and a parameter uncertainty that was 10%–70% higher. The schemes from 4 to 15 s were generally not significantly different in regards to bias and SD. Typically, the reconstruction method 3DRP yielded less frame-sampling dependence and less uncertain results, compared with OSEM, but was on average more biased. Conclusion: Of the 6 sampling schemes investigated in this study, an early frame duration of 6–15 s generally kept both bias and uncertainty to a minimum, for both 3DRP and OSEM reconstructions. Very-short frames of 1 s should be avoided because they typically resulted in the largest parameter bias and uncertainty. Furthermore, 3DRP may be preferred over OSEM for short frames with poor statistics.