RT Journal Article SR Electronic T1 Discrepancies in Dose Calibrator Assays for Various Forms of Therapeutic Iodine-131 JF Journal of Nuclear Medicine Technology JO J. Nucl. Med. Technol. FD Society of Nuclear Medicine SP 84 OP 86 VO 12 IS 2 A1 Levine, Geoffrey A1 Malhi, Balwinder A1 Struttman, Lloyd YR 1984 UL http://tech.snmjournals.org/content/12/2/84.abstract AB Differences in product packaging have the potential to yield differences in assays for the same quantity of radioactivity. We report on the assay variance between packaging of sodium iodide I-131 in capsules and in solutions. We have observed differences of sufficient magnitude that, under certain circumstances, could result in reports of misadministrations to the Nuclear Regulatory Commission when in fact, none has occurred. Establishment of appropriate calibration factors and development of separate dose calibrator potentiometer settings for liquid and capsule dosage forms can minimize the problem.