PT - JOURNAL ARTICLE AU - Ratzlaff, Neil W. AU - Brown, Paul H. AU - Cox, Deborah J. AU - Krishnamurthy, G.T. TI - The Effects of Collimator Malpositioning in Seven Pinhole Tomography DP - 1984 Jun 01 TA - Journal of Nuclear Medicine Technology PG - 51--55 VI - 12 IP - 2 4099 - http://tech.snmjournals.org/content/12/2/51.short 4100 - http://tech.snmjournals.org/content/12/2/51.full SO - J. Nucl. Med. Technol.1984 Jun 01; 12 AB - We tested the effect of seven pinhole collimator positioning on circumferential count profiles during thallium-201 myocardial imaging by imaging a left ventricular phantom at various angles. The seven pinhole raw data images and profiles were identifiable as malpositioned when the collimator was malpositioned by 10° or more. The phantom’s “cold” spot was shown to propagate into adjacent planes and also into other regions of the heart as a function of collimator positioning. We then developed a method for systematically positioning the collimator correctly, which should aid in preventing similar artifacts in clinical studies.