PT - JOURNAL ARTICLE AU - Chen, Qin-sheng AU - Franken, Philippe R. AU - Defrise, Michel AU - Jonckheer, Marc H. AU - Deconinck, Frank TI - Detection and Correction of Patient Motion in SPECT Imaging DP - 1993 Dec 01 TA - Journal of Nuclear Medicine Technology PG - 198--205 VI - 21 IP - 4 4099 - http://tech.snmjournals.org/content/21/4/198.short 4100 - http://tech.snmjournals.org/content/21/4/198.full SO - J. Nucl. Med. Technol.1993 Dec 01; 21 AB - We have developed an approach to detect and correct patient motion that occurs during SPECT imaging. The detection of patient motion is based on frame-to-frame matching, using phase-only matched filtering (POMF). The correction procedure has been designed to correct any translational patient motion and alert the operator to other kinds of patient movement. The precision in correcting translational motion is less than 1 pixel in the axial direction and less than 2 pixels in the lateral direction. Experiments on phantom and patient data have demonstrated that our approach is a reliable, accurate, and sensitive method for detecting and correcting patient motion in SPECT imaging.