RT Journal Article SR Electronic T1 Quality Control Testing of Dose Calibrators JF Journal of Nuclear Medicine Technology JO J. Nucl. Med. Technol. FD Society of Nuclear Medicine SP 124 OP 128 VO 24 IS 2 A1 Robert Y.L. Chu A1 William E. Simon YR 1996 UL http://tech.snmjournals.org/content/24/2/124.abstract AB We have reviewed quality control testing of dose calibrators and have analyzed the potential for error. The uncertainty in half-lives of reference sources can lead to inaccurate assessment of the accuracy and constancy in dose calibrator performance. Molybdenum-99 contamination of 99mTc used as a reference source can appear as a nonlinear response of the dose calibrator. The useful life of a reference source and the tolerable amount of 99Mo contamination are examined.