RT Journal Article SR Electronic T1 Computerized Monitoring of Scintillation Camera Uniformity JF Journal of Nuclear Medicine Technology JO J. Nucl. Med. Technol. FD Society of Nuclear Medicine SP 57 OP 60 VO 16 IS 2 A1 Raff, Ulrich A1 Spitzer, Victor M. YR 1988 UL http://tech.snmjournals.org/content/16/2/57.abstract AB An efficient and practical procedure is outlined to objectively analyze and monitor scintillation camera uniformity performance on a daily basis using a commercially available nuclear medicine computer system. Our methods are illustrated with the four scintillation cameras currently used in our department. Intrinsic camera uniformity is analized daily according to NEMA specifications, and quantitative results are graphed from the most recent 120 calibration days.