RT Journal Article SR Electronic T1 A Method to Assess Collimator Integrity JF Journal of Nuclear Medicine Technology JO J. Nucl. Med. Technol. FD Society of Nuclear Medicine SP 173 OP 175 VO 23 IS 3 A1 Beyea, Don A1 Boudreau, Debbie YR 1995 UL http://tech.snmjournals.org/content/23/3/173.abstract AB Objective: In this paper we evaluate a new method for assessing the structural integrity of parallel-hole collimators. Methods: Images of various collimators were acquired using a dual-energy, x-ray bone densitometry unit. Results: X-ray images of collimators acquired by this method were of high quality and revealed delaminations, hole misalignments, denting and other damage or structural collimator deformities. Conclusion: The bone densitometry unit x-ray beam is parallel, while the beam from a conventional x-ray unit is divergent. X-ray imaging of collimators with a parallel-beam bone densitometry unit is a convenient, inexpensive and technically superior means of assessing collimator integrity.