PT - JOURNAL ARTICLE AU - Beyea, Don AU - Boudreau, Debbie TI - A Method to Assess Collimator Integrity DP - 1995 Sep 01 TA - Journal of Nuclear Medicine Technology PG - 173--175 VI - 23 IP - 3 4099 - http://tech.snmjournals.org/content/23/3/173.short 4100 - http://tech.snmjournals.org/content/23/3/173.full SO - J. Nucl. Med. Technol.1995 Sep 01; 23 AB - Objective: In this paper we evaluate a new method for assessing the structural integrity of parallel-hole collimators. Methods: Images of various collimators were acquired using a dual-energy, x-ray bone densitometry unit. Results: X-ray images of collimators acquired by this method were of high quality and revealed delaminations, hole misalignments, denting and other damage or structural collimator deformities. Conclusion: The bone densitometry unit x-ray beam is parallel, while the beam from a conventional x-ray unit is divergent. X-ray imaging of collimators with a parallel-beam bone densitometry unit is a convenient, inexpensive and technically superior means of assessing collimator integrity.