RT Journal Article SR Electronic T1 Thallium-201 Myocardial Imaging: Comparison of the High Sensitivity and High Resolution Collimators JF Journal of Nuclear Medicine Technology JO J. Nucl. Med. Technol. FD Society of Nuclear Medicine SP 35 OP 36 VO 9 IS 1 A1 Wynant, Gordon E. YR 1981 UL http://tech.snmjournals.org/content/9/1/35.abstract AB Thallium-201 myocardial images of 20 patients undergoing treadmill exercise were evaluated as to the ability of the high sensitivity collimator (HSC) to reveal perfusion defects compared to images obtained using the high resolution collimator (HRC). We acquired 200,000 count images in the anterior, 45° left anterior oblique, and left lateral projections. Average acquisition times were: HSC 3.41 min. and HRC 6.08 min. Fifteen patients who had clearly visible perfusion defects using the HRC showed degraded definition or appreciable loss of definition of the same defects when using the HSC.