PT - JOURNAL ARTICLE AU - McLean, J. R. AU - Wise, Paul TI - Impurities in a <sup>99m</sup>Tc Lung Imaging Kit DP - 1977 Mar 01 TA - Journal of Nuclear Medicine Technology PG - 28--31 VI - 5 IP - 1 4099 - http://tech.snmjournals.org/content/5/1/28.short 4100 - http://tech.snmjournals.org/content/5/1/28.full SO - J. Nucl. Med. Technol.1977 Mar 01; 5 AB - We investigated the nature and significance of three radiochemical impurities that were found in a lung imaging kit using 99mTc-tin-macroaggregated albumin. The most abundant radiochemical impurity was a soluble 99mTc-labeled protein. Free pertechnetate (99mTcO4–) and reduced and hydrolyzed technetium (99mTcO2) were present in relatively low radioconcentrations. The presence of a soluble 99mTc-labeled protein in the preparation at the time of injection modified the biologic performance of the product. We suggest that, to be meaningful, a quality control program must employ assay procedures which are capable of identifying and quantitating all the significant radiochemical impurities that are present in a preparation.