PT - JOURNAL ARTICLE AU - Hiraki, Hitoshi AU - Ito, Toshimune AU - Onoguchi, Masahisa AU - Tsuchikame, Hirotatsu AU - Shishido, Masaaki AU - Maeno, Takafumi AU - Shibutani, Takayuki AU - Sanada, Hiroki TI - Evaluation of Collimators in a High-Resolution, Whole-Body SPECT/CT Device with a Dual-Head Cadmium–Zinc–Telluride Detector for <sup>123</sup>I-FP-CIT SPECT AID - 10.2967/jnmt.122.265328 DP - 2023 Sep 01 TA - Journal of Nuclear Medicine Technology PG - 227--234 VI - 51 IP - 3 4099 - http://tech.snmjournals.org/content/51/3/227.short 4100 - http://tech.snmjournals.org/content/51/3/227.full SO - J. Nucl. Med. Technol.2023 Sep 01; 51 AB - The study aim was to evaluate the adaptation of collimators to 123I-N-fluoropropyl-2b-carbomethoxy-3b-(4-iodophenyl)nortropane (123I-FP-CIT) dopamine transporter SPECT (DAT-SPECT) by a high-resolution whole-body SPECT/CT system with a cadmium–zinc–telluride detector (C-SPECT) in terms of image quality, quantitation, diagnostic performance, and acquisition time. Methods: Using a C-SPECT device equipped with a wide-energy, high-resolution collimator and a medium-energy, high-resolution sensitivity (MEHRS) collimator, we evaluated the image quality and quantification of DAT-SPECT for an anthropomorphic striatal phantom. Ordered-subset expectation maximization iterative reconstruction with resolution recovery, scatter, and attenuation correction was used, and the optimal collimator was determined on the basis of the contrast-to-noise ratio (CNR), percentage contrast, and specific binding ratio. The acquisition time that could be reduced using the optimal collimator was determined. The optimal collimator was used to retrospectively evaluate diagnostic accuracy via receiver-operating-characteristic analysis and specific binding ratios for 41 consecutive patients who underwent DAT-SPECT. Results: When the collimators were compared in the phantom verification, the CNR and percentage contrast were significantly higher for the MEHRS collimator than for the wide-energy high-resolution collimator (P &lt; 0.05). There was no significant difference in the CNR between 30 and 15 min of imaging time using the MEHRS collimator. In the clinical study, the areas under the curve for acquisition times of 30 and 15 min were 0.927 and 0.906, respectively, and the diagnostic accuracies of the DAT-SPECT images did not significantly differ between the 2 times. Conclusion: The MEHRS collimator provided the best results for DAT-SPECT with C-SPECT; shorter acquisition times (&lt;15 min) may be possible with injected activity of 167–186 MBq.