PT - JOURNAL ARTICLE AU - Skweres, Justin AU - Yang, Zhiyun AU - Gonzalez-Toledo, Eduardo TI - Understanding the Cause of an Unreadable Nuclear Medicine Image: A Case of Unexpected Results with <sup>123</sup>I Whole-Body Scintigraphy AID - 10.2967/jnmt.113.136523 DP - 2014 Dec 01 TA - Journal of Nuclear Medicine Technology PG - 302--303 VI - 42 IP - 4 4099 - http://tech.snmjournals.org/content/42/4/302.short 4100 - http://tech.snmjournals.org/content/42/4/302.full SO - J. Nucl. Med. Technol.2014 Dec 01; 42 AB - When unexpected results are obtained with standard image collection, the nuclear medicine physician must consider many technical factors that may have contributed. When image quality is poor, prior radiotracer administration, among other things, should always be considered. Our case demonstrates how knowledge of patient history and basic principles of nuclear medicine physics allows recognition of the septal penetration artifact. This allows the nuclear medicine physician to tailor the exam to an individual patient and obtain the most useful diagnostic information for the clinician.