Abstract
The challenges of hybridizing PET and MRI as a simultaneous modality have been largely overcome in recent times. PET hybridized with MRI has seen the emergence of PET/MRI systems in the clinical setting, and with it comes a responsibility to adapt appropriate facility design, safety practices, protocols and procedures, and clinical opportunity. This article provides an insight into the considerations and challenges associated with PET/MR technology. Given that the nature of PET is well established among the readership of this journal, the article provides an introduction to the foundations of MRI instrumentation and emphasis on specific technologic aspects of PET/MR systems. This article is the second in a 4-part integrated series sponsored by the PET/MR and Publication Committees of the Society of Nuclear Medicine and Molecular Imaging–Technologist Section, building on the previous article (part 1), which was on establishing a facility. In subsequent parts, PET/MRI will be explored on the basis of protocols and procedures (part 3) and applications and clinical cases (part 4).
Although the merits of SPECT, PET, CT and MRI have been widely recognized and reported in both research and clinical practice, the synergistic hybridization of imaging modalities has revolutionized diagnostic imaging. SPECT/CT and PET/CT are in widespread practice, merging the advantages of the high-resolution anatomic information of the CT with the physiologic and molecule-level functional information from the nuclear medicine modalities. More recently, PET/MRI hybrid systems have emerged, although PET/MR was devised and patented in the 1990s (1). The first integrated PET/MRI systems emerged in 2010, and by 2015 there were about 110 PET/MRI systems globally (∼160 in 2020), with Siemens representing nearly 80% of the market, ahead of Philips and GE Healthcare (2). For clarification and consistency, the terms PET/MR and PET/MRI are not used interchangeably in this article. PET/MR is used in reference to the technology and principles, whereas PET/MRI refers to the application of that technology in imaging. The use of MRI versus MR throughout should be self-explanatory. Furthermore, the term imaging system is used in preference to the more archaic use of the term scanner.
A working knowledge of PET is assumed in this discussion but can be refreshed in the previously published work of Turkington (3). The general principles of MRI are outlined below. This is also not a forum for debating the relative merits of PET/MRI against PET/CT. Nonetheless, it would be remiss not to highlight the motivations for PET/MRI (1,4–7). First, PET/MRI produces a reduced patient radiation dose. This is particularly important in pediatric patients, women of childbearing age, and the numerous patients who undergo sequential scanning to monitor response to therapy or recurrence. Compared with PET/CT, PET/MRI can reduce the dose to just 25%, given that 75% of the patient radiation dose is associated with CT. An example would be the reduction in a pediatric whole-body PET/CT scan from 19.5 to 4.6 mSv for PET/MRI. Second, soft-tissue visualization and contrast on MRI are superior to those on CT, enhancing lesion detection and diagnostic accuracy in numerous conditions. MRI adds important complementary functional information to the PET scan, further enhancing outcomes. The opportunity from PET/MRI for multiparametric imaging increases radiomic feature extraction and diagnostic utility. Third, PET/MR offers genuinely simultaneous imaging, which has a positive impact on coregistration, especially in organs that move (e.g., heart and lungs), in lesions that experience physiologic motion (e.g., lung tumors), in organs that change (e.g., filling of the urinary bladder), or in patients who move during scanning (e.g., from pain or anxiety). The simultaneous imaging also allows better synchronization of respiratory and cardiac gating between PET and MRI data, as well as allowing application of motion correction to the PET data. The correction for motion increases lesion detection and quantitative accuracy. Fourth, there is also a reported efficiency when PET and MRI are required, because they are acquired simultaneously. In the absence of PET/MRI, PET/CT is sequential and then requires the additional time of the stand-alone MRI. Finally, image reconstruction offers the chance to correct partial-volume artifacts on the PET images using the MRI coregistration. The net impact of this approach is resolution recovery or enhanced spatial resolution of the PET data.
Despite these advantages, replacing PET/CT with PET/MR is associated with several challenges, such as detector function in a magnetic field, attenuation correction, scatter correction, artifacts, and truncation of the field of view. These challenges are especially true when one is considering truly integrated PET/MR with simultaneous acquisition (1,4,5,7).
FOUNDATIONS OF MRI
For the MRI novice, the basic foundations allow a working knowledge of the principles of MRI. For some, the superficial introduction outlined below will provide sufficient insight. Others will yearn for deeper understanding beyond the scope and word limit of this article. Such insight can be sourced from several key textbooks (8,9). Rather than exploring Newtonian and quantum physics, a simpler model of MRI is offered.
MR Signal
Some nuclei spin on their own axis, which allows them to be thought of as small magnets. Hydrogen is one such nuclei—indeed, the one that produces the strongest magnet. Consider the distribution of hydrogen in the human body. It is clear that hydrogen imaging would be both biologically useful for some tissues and sufficiently abundant in distribution to allow quality imaging. The dispersion of hydrogen in objects is random and fairly uniform, which means the small magnets in opposite alignments cancel one another out to produce a net zero magnetic vector (Fig. 1A). If a magnetic field is applied to the object, then the hydrogen atoms align parallel with the magnetic field and the object has a net magnetic vector; it has become polarized (Fig. 1B) (8–10). In reality, not all hydrogen atoms (let us refer to them as proton dipoles) become aligned with the magnetic field. Approximately 1 per million of proton dipoles realign with the magnetic field, but this is sufficient to create polarization (8). Although a simple schematic representation such as Figure 1 suggests that the atoms have aligned uniformly, in reality they wobble. The wobble is known as precession (Fig. 1C), and the frequency of the precession is called the Larmor frequency and depends on the strength of the magnetic field (8–10). For hydrogen, a 1.5-T MRI system produces 63.9 MHz of precession whereas 3 T produces 127.8 MHz of precession (8,9). Indeed, this might be better expressed as a gyromagnetic ratio of MHz of precession per T. This is important because each dipole nucleus has a unique gyromagnetic ratio. For hydrogen (1H) it is 42.6 MHz/T, whereas others include 19F at 40.1 MHz/T, 31P at 17.2 MHz/T, 23Na at 11.3 MHz/T, and 13C at 10.7 MHz/T.
Excitation
Consider the patient in an MRI system. In the presence of the magnetic field, proton dipoles align with the magnetic field to produce a net magnetic vector (Fig. 1B). This is also referred to as the equilibrium magnetization state. The magnetic field causes each dipole to randomly precess (Fig. 2A). If a pulse of radiofrequency is emitted at the Larmor frequency, it will change the energy state of the hydrogen nuclei (excitation) (8–10). This is achieved using the radiofrequency coil, which pulses at resonance and causes the net magnetic vector to flip to negative on the z-axis. Despite this, the precession remains perpendicular to the z-axis, but instead of being random, the precession of each dipole becomes in phase (Fig. 2B) (8–10). This is referred to as phase coherence and generates the optimal MR signal. The net magnetism returns to the equilibrium magnetism state (relaxation) (8–10). This creates a moving magnetic field, and electric current can then be generated in a radiofrequency receiver coil (Fig. 2C) (8–10). The time for the signal to return to equilibrium is called relaxation time, and the signal produced is referred to as free induction decay (FID) (8–10).
Relaxation
Relaxation time is associated with 2 independent processes, T1 and T2, that occur simultaneously (8–10). T1 relaxation is also known as longitudinal relaxation because it describes the rate at which the magnetism returns to equilibrium. T2 relaxation is also known as transverse relaxation because it describes the rate at which phase-coherent precession dephases. T1 is the time required to return to 63% of the equilibrium value after exposure to a 90° radiofrequency pulse (8–10). Conversely, T2 is the time required to dephase the signal to 37% of its original value. The role of contrast medium is to alter the relaxation times (Fig. 3) (8–10). Relaxivity measures the degree to which a given amount of contrast agent shortens T1 or T2, which means higher relaxivity produces better enhancement (11).
Image Formation
The FID signal intensity decreases over time as a harmonic oscillation (Fig. 4A) (8). Figure 4A represents signal intensity over time, but this can be converted to the frequency domain via Fourier transformation to produce a nuclear MR spectrum (Fig. 4B). The Fourier transformation of the measured MRI signal is referred to as k-space. The clusters of signals at different frequencies represent the specific nuclear MR characteristics of different tissues (Fig. 4C) (8–10). Although this provides an insight into the proportion of various tissues in a sample (as used in nuclear MR spectroscopy), it does not provide information about the location of those tissues. Superimposed on the static magnetic field, a gradient magnetic field is applied with different field strengths, and this allows tissue localization (Fig. 4D) (8–10). A far more complex FID is produced by data collected at multiple projections, which can then be reconstructed using tomographic principles previously described (12).
Pulse Sequences
Although this provides a simple explanation of the MRI process, in reality, FID is not a simple imaging option because of the weak signal (8–10). A 1-pulse sequence would need to be repeated several hundreds of times to generate an adequate signal (8,9). The time between each radiofrequency pulse is called the repetition time (TR), and if sufficient time elapses between each radiofrequency pulse, equilibrium allows the amplitude of successive FIDs to be equal (8,9). A short TR means equilibrium has not been reached and successive FIDs will have smaller amplitudes (partial saturation) (8,9).
Alternatively, multiple pulse sequences can be used. If the 90° radiofrequency pulse is followed by a 180° radiofrequency pulse, an echo signal of the FID can be generated by the rephasing (8,9). The echo signal is termed spin echo and is stronger than the FID signal; it is measured at the peak echo time (8,9). Echo time after the 180° radiofrequency pulse is equal to the time between the 90° radiofrequency pulse and the 180° radiofrequency pulse (8,9). A multiecho spin-echo pulse sequence could be used in which a subsequent 180° radiofrequency pulse follows the previous one, producing another spin echo (8,9). An image can be generated from each of the spin echoes (primary and secondary) at the respective echo time to produce images of differing characteristics. The inversion-recovery pulse sequence starts with a 180° radiofrequency pulse and follows it with a 90° radiofrequency pulse that inverts the net magnetism (8,9). There are a large variety of sequences used in MRI, and the details are beyond the scope of this article; however, those interested are encouraged to do further research on gradient echo sequences, dual-echo sequences, fluid-attenuated inversion-recovery sequences, diffusion-weighted sequences, short-τ inversion recovery sequences, and Dixon chemical shifts.
PET/MR TECHNOLOGY
System Design
One of the principal barriers to hybrid PET/MRI systems is the incompatibility of the technology. The traditional PET system design using photomultiplier tubes (PMTs) is susceptible to even small magnetic fields and, thus, incompatible with the large magnetic fields associated with MRI (1,4–7). Furthermore, the electronics associated with signal processing in PET can create interference in the MR signal (4–7). This limitation gives rise to several solutions or configurations. The first is to have separate PET and MRI systems linked by a single patient table, with each gantry positioned and shielded sufficiently to not interfere with the other (e.g., Philips Ingenuity TF) (Fig. 5A). Alternatively, the detectors themselves could reside within the MRI system and be coupled by optical fibers to PMTs residing outside the magnetic field (Fig. 5B). This design may adopt a split-magnet design allowing the optical fibers to be shorter and the magnet bore wider. The genuinely integrated option would modify detector composition to eliminate susceptibility from and to the magnetic field (e.g., Siemens Biograph mMR, GE Healthcare Signa PET/MR, and United Imaging uPMR 790 PET/MR). Clearly, the integrated system is required to take advantage of simultaneous PET and MRI. One approach is to couple the scintillation detector directly or by optical fibers to avalanche photodiodes (APDs) in radiofrequency shielding (Fig. 5C). Alternatively, silicon photomultiplier (SiPM) detectors can be coupled to the scintillation detector (Fig. 5D).
Commercial Systems
There are 4 commercially available clinical PET/MRI systems on the market. The Signa PET/MR is an integrated system with a 3-T MR component with PET (time of flight) that uses SiPM detectors directly coupled to lutetium-based crystals. The GE Healthcare MR750w wide-bore (70 cm) MRI system can be retrofitted with an SiPM PET insert to produce a 6-cm-bore PET/MR device. The Biograph mMR PET/MR is an integrated system with a 3-T MR component with PET that uses APD detectors and lutetium oxyorthosilicate crystals. The uPMR 790 PET/MR is an integrated system with a 3-T MR component with PET that uses SiPM detectors directly coupled to lutetium yttrium oxyorthosilicate (LYSO) crystals. The Ingenuity TF PET/MR is a sequential system with A 3-T MR component colocated and sharing an imaging system bed with a LYSO and PMT-based PET system.
Detectors
The traditional PET PMT was based on the coupling of bismuth germinate oxide scintillation crystal detectors to PMTs. More recently, the detectors have been upgraded to lutetium-based scintillators such as lutetium oxyorthosilicate and LYSO. Lutetium oxyorthosilicate and LYSO have faster scintillation decay, which makes them more effective in high-count-rate conditions, in additional lines of response for 3-dimensional scanning, in longer axial fields of view, and in time of flight (1,4,6,13). Lutetium oxyorthosilicate and LYSO have typically been coupled to PMTs. PMTs offer signal amplification but are susceptible to magnetic fields and are bulky, both being significant issues for integrating PET into an MRI gantry. APDs are silicon semiconductors that not only are able to operate within the magnetic field but also are compact (1,4,6,7,13). Unfortunately APDs are thermally sensitive, which requires thermal stabilization. Furthermore, the signal gain is very low compared with a traditional PMT, which requires additional electronic signal amplification (1,4,7,13). Both these requirements have been met successfully in traditional sequential PET/CT imaging system design and with some difficulties in PET/MR design. For example, the additional electronics for signal boosting is MR-sensitive, the thermal control creates bulk, and the low signal gain prohibits time of flight.
SiPMs are an alternative to the PMT and APDs, although technically, they are a type of APD (1,7,13). SiPMs are a silicon semiconductor detector operating in Geiger mode and are also referred to in the literature as solid-state photomultipliers and multipixel photon counters (Fig. 6A) (1,7,13). In the SiPM, the detector matrix comprises thousands of pixels that operate as a photon counter in Geiger mode. The pixels are known as single-photon avalanche diodes, and each individual single-photon avalanche diode can be counted independently and simultaneously with other single-photon avalanche diodes (Fig. 6B) (1,13). This approach results in high resolution, high sensitivity, high signal gain, low noise, fast timing, less magnetic susceptibility, and time-of-flight capability (1,13).
CHALLENGES OF PET/MR
Even with the technical issues addressed, PET/MR comes with several challenges. Physical space, shielding, and planning are a major challenge but have been detailed in the first article in this series (14) and will not be duplicated here. The time cost associated with imaging protocols and procedures will only briefly be discussed below since it will be detailed in part 3 of this series of articles.
Time Cost
Although PET/MRI is considered simultaneous in nature, MRI radiofrequency-pulse sequences tend to be longer than PET bed positions (1). A 3-min bed position, for example, may be insufficient for all MRI sequences to be completed, although recent developments have seen the emergence of 4–5 MRI sequences in a 3-min PET bed position. This discrepancy is exacerbated in whole-body imaging typical of oncology studies. The solution under development is fast MRI protocols. These have shown some early success but have several barriers to overcome, including noise reduction. Nonetheless, multiple sequences for whole-body MRI over 20–40 min have a higher time cost than the multiple-bed-position PET covering the same anatomy in 10–20 min (1). An important consideration in this discussion is the actual applications of PET/MRI, which will be discussed in detail in part 4 of this series. Many of the clinical applications or clinical scenarios in which PET/MRI provides a tangible benefit over PET/CT may require only a single bed position—neurologic PET/MRI, cardiac PET/MRI, or head and neck oncology imaging, for example.
Attenuation Correction
A distinct advantage of PET/CT is the ability to perform accurate attenuation correction using a rapid CT acquisition immediately before or after the PET image is acquired. This was a significant advance over the previous methods using 68Ge/68Ga sources to produce transmission scans before the patient is administered the PET radionuclide dose. PET/MRI has no option for producing a transmission scan for attenuation correction because there is no correlation between the attenuation coefficient and MR signal intensity (1,5,6,15). A variety of approaches have been adopted to create an MR-based proxy for the attenuation map; each continues to be constrained by limitations. The challenge of accurate attenuation mapping using MRI is complicated by additional attenuation associated with MR hardware and coils in the field of view (1,6). Furthermore, attenuating structures on the PET image may be positioned outside the acquisition field; an arms-down chest PET/MRI study, for example, may truncate the MR signal from the arms and, thus, be unable to estimate attenuation from the truncated tissues (1,6). The advantage of PET/MRI attenuation correction over PET/CT, if it can be performed accurately, is the reduced patient radiation dose and the simultaneous acquisition that can overcome mismatches between sequentially performed PET and CT attenuation mapping (1).
Simple segmentation and classification of different tissue types using the T1 signal can be constrained by the lack of bone signal and associated attenuation correction (1,5,7,15). Moreover, some tissues, such as lung, can have different attenuation coefficients from one person to the next (1). Another simple approach may be suitable for tissues with uniform attenuation (e.g., the brain) using estimation methods and an attenuation formula (1). This method was used for brains before PET/CT, but the additional influence of MRI coils requires careful mapping of the coil template in the estimated attenuation corrections (1,15). It is possible to also revert to previous PET system designs in which a rotating rod source (68Ge/68Ga) is used for a transmission scan (1,15). Obviously, this approach adds to the time associated with the transmission scan, produces noisy attenuation maps, and remains inferior to CT-based attenuation correction.
Current approaches to attenuation correction on integrated PET/MRI use a 3-dimensional Dixon approach that provides in-phase and out-of-phase datasets for water and fat (6). In essence, this allows segmentation of air, fat, muscle, and lung tissue for attenuation coefficients but lacks accommodation for bone (6). Bone is classified as soft tissue from an attenuation coefficient perspective in this approach. The underestimation of attenuation associated with this bone nonclassification can produce substantial errors in quantitation of both bone and adjacent soft tissue. This could be overcome with the addition of atlas-based methods, but these have limitations around variations from normal anatomy not uncommon in pathologic patients (15). A different approach incorporates ultrashort-echo signals to delineate bone and add to the standard T1 tissue segmentation. The ultrashort-echo-time approach can also be combined with a Dixon sequence (MRI sequence based on chemical shift designed for fat suppression) to produce classes of very short T2 tissues for segmentation (5,6,15). This approach has particular potential in lung imaging.
An important area of development for PET/MRI is pseudo-CT attenuation maps (Fig. 7). There are several limitations in estimating an attenuation map from MRI for PET/MRI that convolutional neural networks (CNNs) may overcome (16). Deep CNNs can produce an attenuation map that closely models the CT-based grounded truth (17,18). Others have integrated the Dixon method with a CNN to generate pseudo-CT for pelvic PET/MRI, with promising results of less than 2% variation from the CT map (19). More accurate attenuation maps also resulted from the use of deep CNN combined with zero-echo-time Dixon pseudo-CT (20). CNN approaches can produce pseudo-CT attenuation maps from the PET sinogram, and this approach was used successfully (<1% error) on 18F-FDG brain scans (21).
The ability to accurately correct for attenuation is essential for accurate quantitation of PET data (1,6). Indeed, the accuracy of relative quantitation markers such as SUV will be undermined if the integrity of attenuation correction is sacrificed.
Artifacts
Both PET and MRI are vulnerable to artifacts, and this vulnerability can undermine not only image quality but also quantitative accuracy. PET/MRI, combined, adds a layer of complexity to potential artifacts. The most common artifact that raises concern is cross-talk between PET and MRI, but this has been largely overcome by the new detector designed discussed previously. It is possible that mechanical misalignment can cause coregistration inaccuracies (6). Imperfect attenuation correction remains a concern for artifact production, and this can be exacerbated by the administration of contrast by interfering with tissue segmentation (6). MRI is susceptible to signal voids associated with items such as implants, and these may also produce attenuation artifacts on the PET scan (6). As already discussed, artifacts can be produced in the PET data by attenuation from the MRI hardware and coils and by truncation of the field of view.
CONCLUSION
Although hybrid imaging with PET/MR plays a role in several clinical situations, there are several technical considerations required in the design and operation of the imaging systems. Specifically, modifications to conventional imaging systems that accommodate integration of the 2 modalities without image-degrading cross talk require deep understanding before the technology can be adopted. Such understanding will have translational benefits to procedural and clinical applications of PET/MRI. Specific design features of PET/MRI help guide the facility planning detailed in part 1 of this series. The integration of PET/MRI affords the opportunity for enhanced protocols and clinical applications which will be explored in more detail in parts 3 and 4 of this series, respectively. Understanding the principles of PET, MRI, and integrated PET/MRI systems provides the foundation for safe, effective, and optimal adoption of PET/MR technology.
DISCLOSURE
No potential conflict of interest relevant to this article was reported.
Footnotes
CE credit: For CE credit, you can access the test for this article, as well as additional JNMT CE tests, online at https://www.snmmilearningcenter.org. Complete the test online no later than September 2024. Your online test will be scored immediately. You may make 3 attempts to pass the test and must answer 80% of the questions correctly to receive 1.0 CEH (Continuing Education Hour) credit. SNMMI members will have their CEH credit added to their VOICE transcript automatically; nonmembers will be able to print out a CE certificate upon successfully completing the test. The online test is free to SNMMI members; nonmembers must pay $15.00 by credit card when logging onto the website to take the test.
Published online March 15, 2021.
REFERENCES
- Received for publication December 19, 2020.
- Accepted for publication February 11, 2021.