Abstract
PET/CT images are usually obtained in the arms-up position in patients with no head and neck pathology and in the arms-down position to image the head and neck area. The arms usually cause artifacts regardless of up or down positioning. These artifacts include beam hardening, scatter, truncation, and cold areas (cold artifacts) in obese or large patients; motion artifacts; implanted-metal-object artifacts; and artifacts related to radiotracer extravasation at the injection site. In this review article, we will discuss the mechanisms of these artifacts and suggest solutions to reduce or eliminate them, such as reviewing the non–attenuation-corrected PET images, performing extended-field-of-view reconstruction, not applying scatter correction, and using software to correct beam-hardening, scatter, and truncation artifacts. We will present various PET/CT images before and after corrections for such artifacts.
PET/CT images are usually obtained in the arms-up position if there is no head and neck pathology and in the arms-down position to image the brain or the head and neck area. When patients cannot raise their arms, artifacts in the body area may result, particularly if the patients are obese or large. However, whether up or down, the arms can cause artifacts (1). Because the area of interest is usually the body (chest, abdomen, or pelvis), artifacts in arms-down imaging are more of a concern than artifacts in arms-up imaging. These artifacts include beam hardening and scatter artifacts between the right and left arm bones, truncation artifacts, photopenic or cold areas (cold artifacts) in the soft tissues and bones in the abdomen and pelvis in obese or large patients, artifacts related to arm motion, artifacts caused by metal objects in the arms, and artifacts from extravasated activity at the injection site.
BEAM HARDENING AND SCATTER
Beam hardening is commonly seen on CT images. Beam-hardening and scatter artifacts can occur between the right and left arm bones on CT images and may result in inadequate correlation of CT data with PET images, mimic pathologic conditions, or result in inaccurate Hounsfield units if the artifacts are severe (2–4). Although not causing significant image artifacts on PET images, beam hardening, if severe, may cause suboptimal attenuation correction (AC), a reduction in the detectability of small lesions, or inaccurate or suboptimal SUVs (Fig. 1). Beam-hardening and scatter artifacts are produced by different mechanisms but appear as multiple linear bands or streaks between two high-attenuation objects (bone, metal, or barium) on CT images (Fig. 1) (3). When the x-ray beam passes through the body, low-energy x-ray photons are attenuated more easily than high-energy photons, which is a particular problem with high atomic number objects (bone, iodine, or metal) (5). The effects of beam hardening on the image can be reduced using iterative reconstruction or with beam-hardening and scatter-correction software (5,6).
TRUNCATION AND COLD ARTIFACTS
Obtaining images in the arms-down position in obese or large patients usually causes truncation artifacts (7). The fact that the CT field of view (FOV) is smaller than the PET FOV in large patients (object extending beyond the scanning field of view on CT) may cause inaccurate AC (8,9). In the arms-down position, the body of the patient extends beyond the transverse CT FOV and creates truncation that is extensive in obese or large patients (1).
In truncation artifacts, at the periphery of the images there is non-AC PET data without corresponding CT images (2,7). To minimize truncation artifacts in the body, patients can be imaged with their arms up. A truncation-correction algorithm or an extended-FOV reconstruction (extending the reconstruction field of view) can correct these artifacts (2,9–11). Newer scanners with larger CT FOVs can reduce truncation artifacts.
The cold artifacts seen on AC PET images (in the abdominal and pelvic regions in obese or large patients imaged in the arms-down position or in the head and neck region in patients imaged in the hyperflexed arms-up position) are due to extensive truncation or are the result of scatter correction. These cold artifacts are not seen on non-AC PET images, after extended FOV reconstruction, after removing the arms from the imaging FOV, or when not applying scatter correction (Figs. 2–6) (1,12). Cold artifacts seen on only one side of the body are likely due to noncentral placement of the patient on the imaging table or to flexing of one arm more than the other, causing unilateral truncation (Figs. 2–4). Keeping the arms straight during the image acquisition and placing the patient symmetrically on the table can reduce these artifacts. Non-AC PET images can grossly assess the areas but usually provide inadequate information in obese or large patients, particularly in deep tissues. When extended-FOV reconstruction is available, it successfully eliminates these cold areas. Scatter correction may induce cold artifacts when it is applied with an iterative reconstruction (scatter-correction artifact), particularly when the patients are scanned with their arms down and the hands and fingers away from the body, resulting in photopenic or cold areas at the level of the hands (12). Changing the type of scatter correction (from relative to absolute) or not using scatter correction can eliminate these artifacts (12–14). Scatter-correction artifacts can also be seen around the kidneys and bladder as photopenic areas (a halo). There are some inaccurate scatter-correction methods in current use in clinical PET imaging that tend to overestimate the scatter contribution (14).
OTHER ARTIFACTS RELATED TO THE ARMS
The other artifacts caused by the arms are related to arm motion, metal objects in the arms, and extravasated activity at the injection site.
Arm motion can cause cold artifacts (areas of washed-out activity) and distorted arm anatomy on PET images (Fig. 7A) (15,16). Such artifacts can affect visual and quantitative assessment of PET uptake. Reviewing the non-AC PET images can help if the motion occurred after the CT acquisition and was not continuous during the PET acquisition. Scatter limitation (no scatter correction) can correct overestimation of scatter caused by arm motion (15,16).
Metal objects in the arms can cause beam-hardening and scatter artifacts. Metal-artifact reduction and metal-deletion techniques can reduce these artifacts (3).
Radiotracer extravasation at the injection site causes various artifacts, such as reconstruction and scatter-correction artifacts that may obscure the adjacent tissues, a hot spot that may mimic a lesion at the injection site, lymph node uptake that may be mistaken for metastasis, and an erroneously low SUV and low uptake in tissues and lesions if extravasation is high (Fig. 5). Accidental intraarterial injection of radiotracer can cause diffusely increased uptake in the tissues distal to the injection site (glove pattern) and may result in underestimation of lesion SUV and inadequate assessment of the arm (Fig. 7B) (17). Performing proper injection technique, removing the arm from the imaging FOV, or using iterative reconstruction can help to reduce or eliminate these artifacts.
CONCLUSION
For accurate interpretation of PET/CT images, it is important to understand and detect certain artifacts. In this review article, we have summarized the PET/CT artifacts caused by the arms-down and arms-up positions and suggested solutions to reduce or avoid these artifacts.
DISCLOSURE
No potential conflict of interest relevant to this article was reported.
Footnotes
Published online Jul. 24, 2020.
CE credit: For CE credit, you can access the test for this article, as well as additional JNMT CE tests, online at https://www.snmmilearningcenter.org. Complete the test online no later than March 2024. Your online test will be scored immediately. You may make 3 attempts to pass the test and must answer 80% of the questions correctly to receive 1.0 CEH (Continuing Education Hour) credit. SNMMI members will have their CEH credit added to their VOICE transcript automatically; nonmembers will be able to print out a CE certificate upon successfully completing the test. The online test is free to SNMMI members; nonmembers must pay $15.00 by credit card when logging onto the website to take the test.