Abstract
The digital PET/CT scanner with digital photon-counting technique promises a shorter scan time, improved small-lesion detectability, and reduced radiation dose for the PET and CT portions of the exam while improving image quality. Methods: In this single-institution retrospective review study, 84 participants who had undergone PET/CT on both analog and digital scanners were analyzed. The aim was to evaluate the impact of image field of view (FOV) and body mass index (BMI) on the digital compared with the analog PET/CT scanners. The participants were categorized into different groups based on their BMI. Total scan times, 18F-FDG doses, and dose–length products (DLP) were collected and compared. Image quality was also assessed by certified nuclear medicine physicians and graded on a scale from 1 to 5. Results: In the skull–to–mid-thigh FOVs, the digital scanner had a scan time shorter by 37% (P < 0.001), a 18F-FDG dose lower by 16% (P < 0.001), but only an 8% reduction in DLP (P = 0.2). In the head-to-toe FOV cases, the digital scanner showed reductions in scan time (33%; P < 0.001), 18F-FDG dose (13%; P < 0.001), and DLP (19%; P < 0.001). When BMI was accounted for, the digital scanner had a scan time shorter by 33% (P < 0.001), as well as a reduced DLP (P < 0.001) and 18F-FDG dose (P < 0.001), with the most prominent changes being in the overweight and obese participants. Image quality was also improved by the digital scanner, with a score of 4.5, versus 4.0 for the analog scanner. Conclusion: The digital scanner has a shorter scan time and lower DLP, requires a lower 18F-FDG dose, and provides improved image quality when compared with the analog scanner. The most impactful difference in scan time, DLP, and 18F-FDG dose were observed in obese and overweight participants.
PET/CT is one of the most accurate molecular imaging modalities and provides elaborate information at the cellular and molecular levels (1). Despite the impact of PET/CT on staging, restaging, and posttreatment evaluation in oncology, ionizing radiation has always been a concern. Efforts to reduce radiation exposure to both patients and staff have been ongoing in the imaging community (2). Advances in digital PET scanner detectors from conventional photomultiplier tubes to one-to-one solid detectors, and improvements in image reconstruction, are some changes taken toward reducing radiation dose in PET/CT imaging systems (3). The PET/CT scan creates high-resolution images while integrating the anatomic framework from CT with the functional images that come from PET.
Philips digital and analog PET/CT scanners were the focus of this study. Components of the analog PET scanner (Gemini) detectors include crystals that can convert photons to flashes of light but cannot count all individual photons. These scintillation crystals are coupled with multiple photomultiplier tubes that convert the flashes of light to electron signals. These electron signals are then sent to a computer for further processing and image production. The digital PET scanner (Vereos) is equipped with digital solid-state counting detectors that count every individual photon created during a PET scan (4,5). Each detector comprises multiple scintillator elements. The high count-rate ability of solid-state detectors comes from one-to-one coupling between the scintillator elements with light-sensing elements, leading to better spatial resolution and faster image acquisition (4).
To investigate the radiation dose in the CT part of the exam, dose–length product (DLP) is used—the total amount of radiation a person receives during a CT examination. On the digital scanner, DLP is the sum of the radiation dose from scout and slices; on the analog scanner, DLP is the radiation exposure from slices only. The digital scanner has promised a shorter scan time, a lower radiation dose from CT, and a lower 18F-FDG dose (6). Since both the digital and the analog systems are used clinically, the possible impact of image field of view (FOV) and body mass index (BMI) in these 2 systems was considered. The purpose of this study was to compare the scan time and radiation dose associated with the digital and analog PET/CT systems. More effectively, the impact of image FOV and BMI on scan time and radiation dose was considered.
MATERIALS AND METHODS
The study was approved by the university’s Institutional Review Board, and the requirement to obtain informed consent was waived.
Participant Characteristics
This study was a single-institution retrospective study of 84 PET/CT patients in an oncology cohort over a period of 5 y. The participants’ age ranged between 18 and 75 y old, and their BMI ranged between 15 and 68.
In total, 110 participants had undergone both the analog and the digital PET/CT scans using the same injected 18F-FDG dose of 5.18 MBq (0.14 mCi/kg), which was the factory-recommended dose. A maximum dose of 555 MBq (15 mCi) for the analog scanner and a maximum dose of 370 MBq (10 mCi) for the digital scanner were used; these were also factory-recommended by Philips. Given the retrospective nature of the study, the 2 scans could not be done on the same day. The time difference between the analog and the digital scans averaged 11 mo.
To calculate BMI in this study, participants’ weight in kilograms was divided by the square of their height in meters (kg/m2). To factor changes in the 18F-FDG dose based on significant weight changes, participants with greater than a 10% variation in their BMI during their follow-up period (n = 26) were excluded from further analyses. The remaining 84 cases (16 skull to mid thigh and 68 head to toe) were categorized into different groups based on BMI: underweight (<18.9), normal weight (19–24.9), overweight (25–29.9), and obese (>30). All participants were instructed to fast for at least 6 h before the scan. The participants’ blood glucose level was measured before injection of the 18F-FDG dose, with an acceptable level being less than 200 mg/dL. Participants were positioned in a quiet, dimly lit room and kept in a warm, unstimulated condition during their standard 60-min uptake time before imaging.
More recently, we adjusted the 18F-FDG dosing in our center from weight-based to BMI-based and included 20 patients who had a prior scan using weight-based dosing. The patients were divided into 3 groups: BMIs of 25 or less, 26–34, and 35 or more.
Image quality was reviewed and analyzed by 2 board-certified nuclear medicine physicians who did not know the type of scanner used. Image quality for all scans was graded on a scale of 1 to 5 (1 = poor; 5 = excellent). Lesion detectability could not be assessed because of the differences in time between the scans.
Image Data Acquisition
Both PET/CT scanners used in this study are fused with a 64-slice CT scanner (7). The Gemini and the Vereos were used as the analog and digital scanners, respectively. Both were American College of Radiology–certified to ensure accurate analysis. For the Gemini (installed at this site in 2006), the CT image data were reconstructed using filtered backprojection technique, a 512 × 512 matrix, a 600-mm FOV in participants with a BMI of less than 34, and a 700-mm FOV in participants with a BMI of more than 34.
To provide significant improvements in image quality combined with dose reduction capabilities, the Vereos (installed at this site in 2018) uses iDose technique to reconstruct CT images (7). iDose is the fourth generation of advanced iterative reconstruction technique and the latest addition to Philips DoseRight tools (7). The FOV for the different BMIs and the matrix size for the digital scanner were similar to those for the analog scanner. The CT scan was performed in a transaxial FOV with similar parameters for all participants regardless of BMI: 120 kVp, variable mAs in a range of 30–100, an average 15.5-cm axial FOV for the skull to the mid thigh and a 56.2-cm transaxial FOV for the whole body, a 4-mm slice thickness, a 4-mm increment, a pitch of 0.704, a rotation time of 0.5 s, and a collimation of 64 × 0.625. The iDose technique was applied for the digital scanner. The CT studies were obtained with an average time of 34.2 s for the skull to the mid thigh and 54.3 s for the whole body.
Ordered-subset expectation maximization was used in PET image reconstruction for both the digital and the analog scanners (3 iterations with 33 subsets for the analog scanner and 3 iterations with 15 subsets for the digital scanner). The PET parameters included 10 frames for the skull to the mid thigh and a maximum of 18 frames for the whole body. In the analog PET exams, photon counting time per frame varied with BMI: for a BMI of less than 25, 60 s per frame for the skull to the mid thigh (frames 1–10) and 30 s per frame for the lower extremity (frames 11–18); for a BMI of 25–29.9, 90 s per frame from the skull to the mid thigh and 30 s per frame for the lower extremity; for a BMI of 30–35, 120 s per frame from the skull to the mid thigh and 30 s per frame for the lower extremity; for a BMI of more than 35, 180 s per frame from the skull to mid thigh and 60 s per frame for the lower extremity (Table 1).
Two protocols were used in the digital PET exams: a regular-body protocol was used for participants with a BMI of 34 or less, and a large-body protocol was used for participants with a BMI of more than 34. In the regular-body protocol, the photon counting time was 75 s per frame from the skull to the mid thigh (frames 1–10) and 37 s per frame for the lower extremity (frames 11–18). In the large-body protocol, the photon counting time was 105 s per frame at each bed position from the skull to the mid thigh and 45 s at each bed position for the lower extremity (Table 2).
Data Analysis
The comparisons were limited to participants who had identical FOVs that included either the skull to the mid thigh or the head to the toes (whole body). The mean DLP, scan time for both PET and CT exams (both FOVs, skull to mid thigh and whole body), and 18F-FDG dose were collected and compared. For statistical analyses, paired t tests were applied. A P value of less than 0.05 was considered significant. All analyses were performed in Microsoft Excel (version 15).
RESULTS
FOV Effect on Scan Time, 18F-FDG Dose, and DLPs
The differences between analog and digital PET/CT scanners regarding scan time, 18F-FDG dose, and DLP were investigated. These data for scans that had been obtained at our institution between 2012 and 2019 were collected and compared.
The analysis revealed a shorter total scan time (from both CT and PET), a lower 18F-FDG dose, and a lower DLP for the digital scanner than for the analog scanner. In terms of skull to mid-thigh FOVs, there was a significant difference in scan time (P < 0.001; 37% reduction) and 18F-FDG dose (P < 0.001; 16% reduction). The difference in DLPs (P = 0.2; 8% reduction) for the Vereos when compared with the Gemini was not significant (Fig. 1; Table 3). In the head-to-toe (whole-body) FOV cases, the Vereos revealed significant differences in scan time (P < 0.001; 33% reduction), 18F-FDG dose (P < 0.001; 13% reduction), and DLP (P < 0.001; 19% reduction) (Fig. 2; Table 3).
Impact of BMI on Scan Time, 18F-FDG, and DLP
The standard procedure for the administration of an 18F-FDG dose for a PET scan has been based on the weight of the participants (8). In a weight-based dosing system, body habitus is not considered and hence this system could be prone to over- or underestimation of the radiation dose that the participants may receive (9). By means of BMI, however, both weight and height are considered and the role of body habitus becomes more prominent. Therefore, BMI is thought be an effective factor in minimizing scan time, DLP, and 18F-FDG dose.
To address this issue, the participant data were evaluated from 2 scanners (digital vs. analog), and a significant difference was found in total scan time between the Vereos and the Gemini (P < 0.001; 33% reduction; Fig. 3; Table 4). Among the BMI groups, scan time differences were the most prominent in the obese patients and the least prominent in the normal-weight patients. The lower scan times (in the digital scanner than the analog scanner) included 16% in the underweight group (P < 0.004), 8% in the normal-weight group (P = 0.0014), 26% in the overweight group (P < 0.001), and 47% in the obese group (P < 0.001) (Fig. 3; Table 4)
Similarly, a significant difference was observed in the total DLP of all cases between the digital and analog systems (P < 0.001). However, when the cases were split into 4 BMI groups, the decrease in DLP for the digital scanner was significant in the overweight (P < 0.001) and obese (P < 0.001) groups but not in the underweight (P = 0.8) group. In the normal-weight group, this trend was the opposite: the DLP was higher for the digital system than for the analog system (P = 0.08) (Fig. 4; Table 4). Consistently, a significantly lower value for the digital scanner (in comparison to the analog) was observed in total 18F-FDG dose for all participants (P < 0.001), and again, significantly lower values were found for normal-weight (P = 0.012), overweight (P < 0.001), and obese (P < 0.001) participants but not for underweight (P = 0.5) participants. Of note, because the 18F-FDG dose had been administrated per kilogram of body weight, the major difference between the analog and the digital scans was in participants with higher BMIs (overweight and obese), most notably in participants with a BMI of more than 30. This finding suggests that these 2 groups of participants might benefit the most (lower 18F-FDG doses; Fig. 5; Table 4).
These encouraging findings merit further investigation regarding the effect of BMI on administered 18F-FDG dose. Therefore, we changed the administered 18F-FDG dose from weight-based to BMI-based. In this new BMI-based system (20 participants), the administered 18F-FDG dose was 222 MBq (6 mCi) for participants with a BMI of less than 25, 296 MBq (8 mCi) for participants with a BMI of 26–34, and 370 MBq (10 mCi) for participants with a BMI of more than 35 in the digital system. The data on the administered 18F-FDG dose from the Gemini, the Vereos weight-based dose, and the Veroes BMI-based dose were collected and compared with one another. The results show a significant difference between the Gemini weight-based dose and the Vereos weight-based dose (P < 0.001; 34% reduction) and between the Vereos weight-based dose and the Vereos BMI-based dose (P < 0.001; 22% reduction) (Fig. 6; Table 5). However, the new dosing system is an ongoing research project designed to reduce the radiation dose to participants.
Image Quality
Review of the image quality demonstrated a mean score of 4.0 for the analog scanner and 4.5 for the digital scanner. Overall, the images appeared less noisy on the digital scanner than on the analog scanner.
DISCUSSION
Historically, the early idea of PET was developed in 1950 and the first scanner was innovated in 1970 at Washington University in Saint Louis, MO (3). From the early single-slice detector design to the commercial scanners of today, the benefits of PET/CT in oncologic studies cannot be denied, but there have always been aspirations of reducing the 18F-FDG dose and scan time. The single pair of detectors in planar imaging has evolved into the current one-to-one solid detectors in PET, which has led to advanced acquisition electronics, data processing, and image analysis over the past 50 years (3). The latest improvement in PET/CT scanners has been the digital photon-counting technique, which promises a shorter scan time and a lower radiation dose while improving image quality and small-lesion detectability.
Multiple studies (10,11) have reported the superiority of digital PET/CT systems over analog scanners respecting improved small-lesion detectability and image quality (10,11). However, to our knowledge, no study has addressed the differences between digital and analog PET/CT scanners with respect to scan time, radiation dose from administered dose, and radiation dose from the CT portion of the scan. In this study, scan time, 18F-FDG dose, and DLP were evaluated in digital and analog PET/CT scanners, and significant differences in these factors were observed between the digital and analog scanners. In line with expectations, scan time and radiation dose were lower in digital than in analog PET/CT scanners (12). With a few exceptions (underweight patients), results from all patients indicated a significantly shorter scan time and reduced radiation dose in the digital scanner than in the analog scanner (Fig. 3; Table 4).
The DLP in the skull–to–mid-thigh FOV showed no significant difference that could be raised from the differences in calculation methods between digital and analog scanners. The digital DLP is calculated using the sum of the radiation dose from both the scout and the slices. However, for the analog scanner, the DLP could use the sum of the radiation from only the slices since the scout was not available. This in theory should give the digital scanner a higher DLP; however, the new iDose iterative reconstruction technique is able to maintain image quality with a lower radiation dose. Our evaluation likely underestimates the radiation savings since the scout is not included in the analog DLP calculation. In terms of different BMI groups, the DLP and 18F-FDG dose values were not beneficial for underweight participants, for 2 potential reasons. The first is the mentioned reason for variations in DLP calculations between digital and analog scanners. The second comes from undersampling in underweight groups due to exclusion criteria (most of the participants in these groups were excluded from further analyses because there was more than a 10% variation in their BMIs).
Since the administered 18F-FDG dose was weight-based, the results illustrate that reducing the 18F-FDG dose would not be beneficial to underweight participants but would be beneficial to normal-weight, overweight, and obese participants. Per our findings, the dosing system has been changed to a BMI-based system instead of weight-based at our center. In this new BMI-based system (20 participants), the administered 18F-FDG dose is 222 MBq (6 mCi) for participants with a BMI of less than 25, 296 MBq (8 mCi) for participants with a BMI of 26–34, and 370 MBq (10 mCi) for participants with a BMI of more than 35 for the Vereos PET/CT system (Fig. 6; Table 5).
Some limitations should be considered regarding the interpretation and generalization of our findings. First, this was a single-institution study. Second, it is important to remember that in a digital scanner, DLP is not calculated the same way as in an analog scanner but that DLP has been used for evaluation of radiation in CT, and as this study was retrospective, this parameter could not be edited. Third, since this was a retrospective study, our findings need to be reevaluated prospectively. In addition, lesion detectability could not be assessed because of the different time points in the scans of each patient. However, this has been assessed in a recent study in which patients had both scans after a single 18F-FDG injection and confirmed that the digital scanner demonstrated superior small-lesion detection (13). Fourth, we compared digital and analog PET/CT scanners from the same manufacturer. Future research could compare data acquired from a digital PET/CT scanner from multiple manufacturers, again comparing the scan time, 18F-FDG dose, and DLP. Finally, we did not find beneficial effects for underweight participants, and it is worth noting that 2 potential intrinsic factors could be involved here: the first is that in this short period between the first and follow-up scans for each patient, a big shift in the weight of each participant would be unlikely; the second is that, yet, these small changes would significantly affect the “percentage of weight change” in the underweight patients but not in the overweight or obese patients. Therefore, undersampling for the underweight patients might be a caveat of our study. Expanding this analysis to a larger population of underweight patients might help clarify this conundrum.
CONCLUSION
Compared with the Gemini analog PET/CT system, the Vereos digital PET/CT scanner provides improved image quality with the benefits of shorter scan time, lower radiation exposure dose, and lower administered 18F-FDG dose, which lead to a lower radiation dose to the technologist and the public. On the basis of this study, we conclude that the digital PET/CT scanner is a beneficial molecular imaging modality in view of its lower radiation dose and shorter scan time.
DISCLOSURE
No potential conflict of interest relevant to this article was reported.