Abstract
We quantitatively and qualitatively investigated 2-dimensional (2D) and 3-dimensional (3D) imaging with scan-time reduction in 14 patients with 17 lesions, who had known or suspected head and neck cancer, using a bismuth germanate (BGO) crystal based PET/CT scanner with noise-matched images. Methods: A 2D and 3D acquisition protocol using scan-time reduction on an axial single field of view resulted in a 2D 4-, 3D 4-, 3D 3-, 2D 3-, 2D 2-, and 3D 2-min scan sequence to minimize redistribution and decay bias. Tumor maximum standardized uptake values (SUVmax) and tumor mean standardized uptake values (SUVmean) were recorded, and two observers in consensus investigated lesion conspicuity between 2D and 3D paired 4-, 3-, and 2-min noise-matched images. Results: We found some minor advantages quantitatively in favor of 2D scanning, with higher mean SUVmax, and qualitatively in favor of 3D scanning, with lesion conspicuity preference. In our cohort, no great advantage or disadvantage to using either acquisition mode was observed, and all lesions were seen irrespective of acquisition mode and scan time. Conclusion: In head and neck cancer patients, we can recommend a scan-time reduction from 4 to 3 min/bed position in 2D acquisitions with a BGO-based PET/CT scanner, using our imaging protocol and reconstruction defaults.
Bismuth germanate (BGO) crystal–based PET and more recent combined PET/CT scanners are in widespread clinical use, predominantly to image 18F-FDG in cancer studies, but there is still controversy over potential advantages or disadvantages of 2-dimensional (2D) versus 3-dimensional (3D) acquisition and scan-time reduction clinically. In 2D scanning, interplane septa reduce image-degrading random and scattered counts at the expense of overall system sensitivity, with the possibility of long scan times to improve count statistics. In 3D scanning, septa are retracted (typically allowing a large increase in such counts), thereby requiring imaging systems with characteristics of low dead time, fast decay crystal or readout electronics, good energy resolution, and scatter correction to optimize the signal-to-noise ratio in images.
Such factors raise questions about the clinical advantages and disadvantages of 2D acquisition, compared with 3D acquisition (1), regarding image quality and potential scan-time and dose reduction. Additional parameters that may influence this comparison are patient demographics, fasting time before scanning, choice and amount of tracer administered, time to start scanning after injection, data representation and reconstruction algorithms used (2), scanner used (3), and type or site of cancer.
Phantom studies have been used to investigate the lesion-to-background ratios for 2D and 3D acquisition modes in whole-body (WB) scanning. In 1 phantom study, injected activity was optimized using 2D and 3D noise equivalent counting (NEC) rates, but no advantage was seen (4). Similarly, no improvement was found between scanning modes in a small tumor or lymph node phantom study (5). However, 3D imaging was favored when the injected activity was optimized for this mode, but 2D imaging was found to be superior for detecting lesions at higher doses (6).
This uncertainty led some (7) to investigate the optimization of the clinical performance of a BGO-based scanner by interleaving 2D (scan duration, 5 min/bed position) acquisitions with 3D (scan duration, 3, 4, and 5 min/bed position) acquisitions in WB clinical acquisitions, with phantom studies used to optimize 2D and 3D imaging parameters.
Matching noise in paired 2D and 3D scans, by the application of postreconstruction filtering, can reduce bias in uptake measurements arising through differences in 2D and 3D reconstruction techniques. A 2D versus 3D simulated WB clinical study was performed with dynamic acquisitions on a lutetium oxyorthosilicate–based scanner using a single field of view (SFOV) (8) with noise-matched images. Transmission scanning was performed using 68Ge rod sources, as in the studies by Visvikis et al. (7) and Lodge et al. (8); each of these publications concluded that 3 dimensions offered clinical image quality comparable to, if not better than, 2 dimensions.
Emphasis has also been placed on 2D versus 3D comparison, with imaging optimization on specific tumor sites rather than general WB scanning. In 1 such brain study, Dhawan et al. (9) used an interleaved 2D and 3D protocol for acquiring sequential scans and found that 3D 18F-FDG studies were comparable to half the injected activity or scan duration of similar 2D scans. However, different conclusions were made in high-counting-rate cardiac studies regarding the advantage, if any (10), of 2D (11) versus 3D acquisition (12). Following recent reports on WB scanning (13), advances in reconstruction algorithms on the latest BGO-based scanners have led to advantages in 3D over 2D acquisition for the imaging of specific lesions. However, some ambiguity exists in the literature regarding the potential clinical benefits of site-dependent 2D versus 3D imaging. This effect may be more pronounced in older BGO systems with moderate crystal energy resolution, relatively long decay times, and less sophisticated image reconstruction algorithm implementation. Our institution has clinical experience in 18F-FDG head and neck cancer studies and research (14) and associated therapies (15). Accordingly, we investigated the advantages and disadvantages of 2D versus 3D acquisitions and scan-time reduction on a PET/CT BGO crystal scanner (Discovery ST; GE Healthcare) using paired 4-, 3-, and 2-min scan times on an SFOV.
Our aim was to investigate lesion maximum standardized uptake values (SUVmax) and lesion mean standardized uptake values (SUVmean) for each respective 2D and 3D scan pair and for each scan time under conditions of matched noise, thus minimizing 2D and 3D reconstruction bias effects. Two experienced observers then investigated the clinical significance of these quantitative results by performing a qualitative visual comparison of lesion conspicuity between paired scans.
MATERIALS AND METHODS
Informed written consent was obtained from 14 patients with known or suspected head and neck cancer (total of 17 lesions) who were recruited into this prospective, single-institution, nonrandomized study approved by the local Mount Vernon Ethics Committee.
Our cohort consisted of 12 men and 2 women (average age ± SD, 51 ± 13 y; average weight ± SD, 75 ± 12 kg). Patients fasted for at least 6 h before an intravenous administration of 18F-FDG (4.5 MBq/kg of body weight). WB scans began on average 65 ± 6 min after the injection of the tracer, and the average 18F-FDG uptake time to the first SFOV scan (i.e., in 2 dimensions for 4 min) was 92 ± 6 min.
The entire scanning protocol comprised WB PET/CT from the base of the brain to the groin. All patients were scanned with their arms down by their sides to reduce noise in head and neck images. The following were typical WB, spiral, 4-slice CT parameters: 0.8-s tube rotation, 140 kVp, 80 mA, and 1.5:1 pitch. Routine 2D WB PET started approximately 60 min after injection, with a duration of 4 min/bed position and an overlap of 3 slices.
Reconstructed PET/CT WB images were used to locate head and neck lesions for calculating bed travel distances, to position the highest lesion uptake region in the PET SFOV center. In this way, bias from nonuniform sensitivity across the 3D FOV for interleaved scans that followed was minimized. After a regular WB scan, the patient continued to lie still as the bed position was translated the required distance; paired, interleaved 2D and 3D PET images were acquired for each scan time (Fig. 1).
Image Reconstruction
The National Electrical Manufacturers Association/International Electrotechnical Commission (NEMA/IEC) 2000 PET phantom, consisting of 6 spheres (10-, 13-, 17-, 22-, 28-, and 37-mm diameters) situated in a background phantom, was used to deduce optimized reconstruction parameters for SFOV 2D and 3D scanning. A clinically realistic hot sphere–to–background activity ratio of approximately 4:1 was used, and reconstruction parameters were considered optimized when the hot sphere SUVmax was close to or had reached a plateau signifying convergence in the reconstruction.
All 2D and 3D SFOV PET transaxial emission images were reconstructed to a 128 × 128 × 47 matrix, with pixel dimensions of 4.68 × 4.68 × 3.27 mm. 2D reconstructions were performed using an ordered-subset expectation maximization (OSEM) iterative algorithm (16) with scatter correction (17). 3D reconstructions used Fourier rebinning, followed by weighted least-squares iterative reconstruction OSEM (18) and model-based scatter correction (19). Additional 2D and 3D corrections were performed for randoms (singles), decay, dead time, geometry, and attenuation.
Image Analysis
For each patient, a predominantly circular region of interest (ROI) was drawn by hand using the scanner manufacturer's imaging workstation–integrated software to encompass the tumor, thus allowing SUVmax and SUVmean to be recorded on each patient's 2D 4-min SFOV axial scan. An ROI was carefully positioned close to the lesion in the same image slice, to acquire a uniform mean background uptake. The percentage coefficient of variation was defined as 100 multiplied by (SD/mean) of the average background ROI counts; this parameter was used as a quantitative measure of image noise. To minimize variability, the same lesion and background ROIs were then applied in the same locations to each of the other scans composed of different acquisition modes and times.
Our study incorporated noise-matched scans to improve efficacy. For every patient, each 2D or 3D SFOV acquisition was reconstructed using a gaussian postreconstruction filter of 1-, 2-, 3-, 4.5-, 5.14-, 6-, 7-, 8-, 9-, and 10-mm full width at half maximum, resulting in 10 different reconstructed scans for each 2D or 3D SFOV acquisition. For each reconstructed scan, a noise value, established close to the lesion SUVmax, was recorded using the same background ROI, resulting in a total of 10 noise values for each 2D or 3D SFOV acquisition. We compared each of these 10 noise values from the 2D acquisition with each of the 10 noise values from the paired 3D acquisition for each SFOV (e.g., 2D 4 min and 3D 4 min) to establish the closest noise match. The 2D and 3D respective lesion SUVmax and SUVmean were then recorded. This process was repeated, ensuring that noise match was specific to each patient's paired 2D and 3D acquisition and specific to each 4-, 3-, and 2-min scan.
For qualitative analysis, each noise-matched pair of 2D and 3D scans acquired for 4, 3, and 2 min was compared for lesion conspicuity by 2 experienced radiologists in joint consensus and rated accordingly with a score of −2, −1, 0, +1, or +2, in preference for 1 of the 2 scans. A score of 0 signified that no preference was expressed for either scan by the observers, a score of −1 or +1 implied that one scan appeared superior to the other, and a score of −2 or +2 suggested that one scan appeared far superior to the other.
Statistical Analysis
SUVmax and SUVmean data acquired in 2 and 3 dimensions for different scan times were positively skewed, requiring log10 transformation to approximate gaussian behavior for statistical analysis. The Shapiro–Wilk test is preferred for investigating normality in small sample size distributions and was used in this study.
Linear regression and the Pearson correlation coefficient were used to investigate the possibility of bias in results introduced through correlations between SUVmax with weight, height, body mass index (BMI), and blood glucose.
Differences between images for recorded SUVmax at each scan time were analyzed using the paired t test; a P value of less than 0.05 was considered statistically significant, and a P value of less than 0.0001 was considered highly statistically significant. The same was also true for SUVmean and noise measurements, respectively, in paired scans.
For qualitative analysis, the scoring system resulted in a preference scale that was parametric, for which a mean and 95% confidence interval (CI) were constructed. A consistent preference for 1 scan in the direction indicated by the coding at a 5% level was suggested when the 95% CI did not cross 0 and was consistent with a 1-sample t test.
RESULTS
2D and 3D Reconstruction Parameters
Figure 2 shows SUVmax normalized to the maximum for each acquisition mode and scan duration using the 28-mm-diameter sphere in the NEMA/IEC 2000 PET phantom and illustrates increased uptake with increasing OSEM number of iterations multiplied by subsets in 2D and 3D image reconstruction. The same trend of hot sphere uptake reaching a plateau with increased number of reconstruction iterations multiplied by subsets was observed with the other 5 spheres in the phantom.
We selected 21 iterations and 30 subsets (iterations × subsets, 630) for optimized 2D reconstruction and 20 iterations and 32 subsets (iterations × subsets, 640) for optimized 3D reconstruction, based on SUVmax from all spheres reaching or being close to the plateau to ensure convergence.
Noise Match in Paired Scans
Figure 3 shows the distribution of noise-matched data between 4-, 3-, and 2-min paired 2D and 3D images for each patient. Results from this noise-matched plot are presented in Table 1, which shows the mean noise value, SD, difference in means, and associated statistical significance for paired scans of different scan durations. The nonsignificant differences observed for 4-, 3-, and 2-min scans are desirable, suggesting a close match in noise characteristics between 2D and 3D paired scans.
SUVmean in Paired Scans
Figure 4 illustrates the distribution in SUVmean data between paired 2D and 3D images for each patient with 4-, 3-, and 2-min scans. Results from this SUVmean plot are presented in Table 2, which shows the mean SUVmean, SD, difference in means, and associated statistical significance for paired scans of different scan durations. Nonsignificant differences are reported for each scan pair.
SUVmax in Paired Scans
Figure 5 depicts the distribution in SUVmax data between paired 2D and 3D images for each patient with 4-, 3-, and 2-min scans. Results from this SUVmax plot are presented in Table 3, which shows the mean SUVmax, SD, difference in means, and associated statistical significance for paired scans of different scan durations. A significant difference for only the paired 2D 4-min and 3D 4-min scans (95% CI, 0.03–0.18; paired t test, t = 2.90, P = 0.01) was observed.
Qualitative Observer Scores
Figure 6 depicts SFOV noise-matched axial slices that contain the SUVmax from 1 patient, acquired with 4-, 3-, and 2-min scans.
The observer scores for preference in paired scans are shown in Table 4.
Qualitative results for observer-scored paired scans using the 1-sample t test are also shown in Table 4. The mean, SE, 95% CI, and statistical significance of the score between paired scans are depicted. Consensus agreement between experienced radiologists resulted in a significant preference for the conspicuity of lesions in paired 3D 3-min scans, compared with 2D 3-min scans, and in paired 3D 4-min scans, compared with 2D 4-min scans, only.
SUVmax Correlations
No statistically significant correlation between age at the time of the scan, weight, height, BMI, and blood glucose with SUVmax was observed for this cohort of patients. Table 5 shows the respective P values calculated for 2D and 3D 4-, 3-, and 2-min scans for these parameters, calculated using linear regression and the Pearson correlation coefficient.
DISCUSSION
In this study, we focused on genuine, consecutive patients with suspected head and neck cancer who had routine clinical WB scans in our center using standard protocols. No attempt was made to confirm diagnosis with histology. Similar to another study (13), our protocol allowed WB scans to be obtained under standard imaging conditions for radiologists; 2D and 3D research SFOV images with reduced scan times were then acquired if the patient agreed to continue. Our protocol provided not only valuable information about the advantages and disadvantages of 2 dimensions versus 3 dimensions with scan-time reduction in head and neck cancer but also a potential clinical benefit of optimized scanning on those occasions in which more head and neck patients require scanning in a restricted time period. This benefit is particularly true in situations in which delays occurred (e.g., because of machine failure or reduced 18F-FDG supply or the like).
The interleaved SFOV scan sequence reduced 2D or 3D measurement bias, and paired scans that were close in time resulted in insignificant tracer concentration redistribution and decay difference effects between acquisition modes. Because interleaved SFOV acquisitions did not require additional CT exposure beyond their initial WB CT used for attenuation correction, the PET/CT protocol minimized each patient's WB effective dose to ionizing radiation. If movement occurred between the WB scan and subsequent 2D and 3D SFOV acquisitions, then misregistration artifacts would likely have influenced both 2D and 3D scans equally. With our protocol, we could not account for movement between paired interleaved 2D and 3D scans. However, all SFOV images were investigated visually for gross misregistration, to minimize attenuation correction errors, and no errors were observed.
To minimize movement artifacts in head and neck cancer patients, routine 2D WB imaging (scan duration, 4 min/bed position) was performed in the cranial to caudal direction, starting on average 65 min after the injection of the tracer. Head and neck lesions were seen during the first or second bed positions during scanning. The first 2D 4-min SFOV scan was obtained after the WB scans, on average 92 min after the injection of the tracer. The average delay between ending the head and neck lesion imaging part of the standard WB scan and starting the first optimized SFOV scan was 21 min. This time delay had the potential to introduce some redistribution and decay bias in uptake between the routine clinical WB and research SFOV scans; however, no significant differences in mean SUVmax or SUVmean were observed (P = 0.28 and P = 0.07, respectively).
We validated the efficacy of data in this study by investigating tumor uptake bias and noted no correlation between SUVmax and age, weight, height, BMI, or blood glucose (Table 5). This result confirmed our expectation, because the average patient BMI ± SD (24.9 ± 1.8) in our cohort did not include obese patients (BMI > 30) in whom one might potentially expect more correlations. Incidentally, in obese subjects 2D imaging may be preferred because of increased photon scatter rejection using a collimator. To further authenticate our methodology and data, we compared like with like when contrasting 2D and 3D scans by noise-matching images between paired scans (8). A close match in noise between paired scans for 2, 3, and 4 min, respectively, was achieved (Fig. 3) before further data analysis.
No significant differences were subsequently seen between mean SUVmean in paired scans (Fig. 4). In addition, we observed no statistically significant differences between any individual mean 2D SUVmean of different scan times: 2D 4-min vs. 2D 3-min scans (P = 0.78), 2D 4-min vs. 2D 2-min scans (P = 0.27), and 2D 3-min vs. 2D 2-min scans (P = 0.32). The same was true for individual mean 3D SUVmean and between all combinations of mean 2D SUVmean, compared with mean 3D SUVmean, with the same or different scan times.
For scan-time optimization in 2D mode quantitatively, we observed no significant difference between any of the mean SUVmax for different time intervals (Fig. 5): 2D 4-min vs. 2D 3-min scans (P = 0.41), 2D 4-min vs. 2D 2-min scans (P = 0.93), or 2D 3-min vs. 2D 2-min scans (P = 0.50). Accordingly, the qualitative observer visual analysis scores reflected this result because no significant preference was found between 2D 4-, 3-, and 2-min scans. Furthermore, the difference in noise means of 1.91% between 2D 4-min scans and 2D 3-min scans was not significant (P = 0.07), whereas the corresponding 2D 4- and 2D 2-min difference of 6.24% was significant (95% CI, −0.31 to −0.01; paired t test, t = −2.26, P = 0.04). All these data support our pragmatic suggestion of a scan-time reduction per bed position in 2 dimensions from 4 to 3 min as routine for head and neck cancer; however, because our sample size and BMI were limited in this study, we would caution against routinely using 2 min/bed position.
In the case of scan-time optimization in 3D mode quantitatively, we saw a significant difference in mean SUVmax between 3D 4-min scans and 3D 2-min scans only (95% CI, −0.14 to −0.02; paired t test, t = −2.88, P = 0.01). Qualitative visual analysis scores for 3D mode showed no significant preferences for any scan time in comparison with another. Again, we can suggest 3D imaging routinely using 3 min/bed position acquisition in head and neck cancer but would urge caution against using 2 min/bed position for the same reasons as declared above in the 2D case.
When comparing 2D with 3D scan modes quantitatively using paired SUVmax data of the same scan duration (Fig. 5), mean 2D SUVmax at 4-, 3-, and 2-min scan durations was greater in magnitude than equivalent mean 3D SUVmax. Nonetheless, only the 4-min difference between the mean 2D SUVmax and mean 3D SUVmax was significant (P = 0.01), suggesting a slight quantitative 2D scanning, compared with 3D scanning, advantage using SUVmax. However, comparison of visual analysis observer scores between 2D mode and 3D mode for the same scan time confirmed that 3 dimensions were significantly preferred over 2 dimensions at 4 (P = 0.0001) and 3 min (P = 0.0001) (Table 4), suggesting an overall qualitative advantage of 3D scans, compared with 2D scans, in head and neck cancer.
The results of the 2D, 3D, and 2D versus 3D scans presented here illustrate the potential for slight advantages in 2D scanning quantitatively and in 3D scanning qualitatively, thus highlighting both the benefits of each mode and the lack of dominance of one mode over the other in head and neck imaging on this BGO system. The lack of significant difference in paired 2D and 3D SUVmean data further supports the equivalence of imaging modes and generally complements the SUVmax data.
However, SUVmax measurements are typically more reproducible than are SUVmean measurements, for which the entire lesion must be accurately delineated by an ROI for accurate values (20) and the influence of background in the ROI influences the result; therefore, we focus on SUVmax for quantitative analysis.
Our results are validated by comparable 2D and 3D outcomes in work performed by other groups. Congruent results of 2D and 3D quantitative image analysis have been presented in phantom studies on the same make and model of BGO scanner as the one used in this study, using widely accepted NEMA tests (21), and on more modern BGO systems (22). 2D and 3D mode equivalence is also true for qualitative analysis in WB clinical studies on an older BGO system (7); in addition, similar to our study, qualitative preference in 3D scans was expressed by physicians using current GE scanners (23) consisting of different crystals.
There were several limitations to this study. We performed our optimization process with the NEMA-recommended phantom using a hot sphere–to–background ratio of approximately 4:1, but there are other ratios of clinical significance (e.g., 2:1 ratio) relating to patient management.
The entire scan protocol was not simple to implement clinically; to reduce complication and possible errors in execution, we did not implement an inversion in the order of SFOV scans in half the patients. Consequently, we may have introduced some bias. However, for 3-min scans, which we recommend, the order of scans was reversed so that 2 dimensions followed 3 dimensions and the difference in mean SUVmax between these was not significant (P = 0.17), implying that scan order did not heavily bias results for our suggested scan duration of 3 min.
To ensure convergence, our SFOV reconstruction parameters were at a higher number of iterations and subsets than standard WB imaging. However, there was no significant difference between head and neck lesion mean SUVmax or mean SUVmean between the standard WB and first SFOV reconstructed scans. Because we obtained the same results when performing the study without noise matching or optimization using standard SFOV reconstructions (i.e., 2 dimensions are quantitatively superior, and 3 dimensions are qualitatively preferred (24)), further validation is unwarranted.
We observed variations in the recorded data similar to those shown by others and believe these differences may be accounted for by a combination of factors. These include calibration errors between 2D scans and 3D scans, combined with differences in scatter-correction accuracy applied between acquisition modes. During image reconstruction, it may be possible that 3D scatter correction overcorrects in some cases, leading to reduced 3D uptake, whereas 2D SUVmax may be overemphasized in noisy backgrounds. Gaussian filter selection for 2D and 3D noise matching may have influenced spatial resolution and also played a role. However, this effect was variable, thereby reducing a potential systematic bias in resolution (25); no significant differences between standard WB and first SFOV lesion SUVmax or SUVmean were recorded.
In addition to these limitations, we advise some caution in applying the outcome of this clinical study to routine WB imaging and to other tumor sites. In this work, we investigated head and neck lesions in the center of the FOV and our protocol did not support the study of varied slice overlaps or other image-reconstruction parameters that influence 2D or 3D imaging. Moreover, none of the patients in this study was clinically obese and, hence, the potential bias for 2D scanning that might be the case in such conditions was unlikely in this study.
In support of our study, other studies that performed work without optimization, for example, those with a form of resolution matching (8) or more recently those without (13), acknowledged this drawback. Taking work performed without optimization into account, along with noise-matched images and the similarity of our results to those of others, this study is valid and overcomes the limitations presented.
3D imaging in this study generally favored lesion conspicuity, but all lesions were clearly seen in all patients regardless. Instead, we propose that 2D imaging and SUVmax quantitation are important parameters for optimization regarding scan time per bed position. Furthermore, no significant differences were seen between 2D 4-min SFOV and 2D 3-min SFOV acquisitions for mean SUVmax, mean SUVmean, or mean noise (P = 0. 41, P = 0.78, and P = 0.07, respectively). Thus, we can recommend 2D 3-min scans as a pragmatic alternative to 2D 4-min acquisitions for reduced scan time per bed position in head and neck cancer.
CONCLUSION
In noise-matched paired scans, we have shown a slight quantitative advantage to 2D scanning, with higher mean SUVmax, and a qualitative advantage to 3D scanning, with preference for lesion conspicuity. However, all lesions were identified in all scans regardless of acquisition mode, and consequently there was no overwhelming advantage or disadvantage to using either acquisition mode. For quantitative studies in head and neck cancer, we recommend routine 2D clinical imaging with a 3-min scan duration over an SFOV using this BGO-based PET/CT scanner with our imaging protocol and reconstruction defaults.