A Rapid Radiochemical Purity Testing Method for 99mTc-Tetrofosmin ================================================================= * Laura A. Eggert * Michael D. Dick * Douglas W. Mahoney * Joseph P. Olson * Gregory L. Werner * Joseph C. Hung ## Abstract The standard radiochemical purity (RCP) testing method for 99mTc-tetrofosmin as described in the package insert requires extensive time (20–30 min) and considerable skill to achieve accurate results. Additionally, the instant thin-layer chromatography strip impregnated with silica gel (2×20 cm) used in the standard method will not be commercially available in the future. The purpose of this study was to evaluate whether a method developed by our laboratory for RCP testing of 99mTc-sestamibi could also be used as an alternative method for the RCP assay of 99mTc-tetrofosmin. **Methods:** The alternative RCP testing system consisted of a precut paper strip (1×8.5 cm) from solvent saturation pads (Pall Corp.) as the stationary phase, with 1:1 chloroform:tetrahydrofuran used as the mobile phase. To validate the reliability of the alternative method, RCP values from 17 kit preparations were compared with the 2 methods. Kits were reconstituted according to the package insert instructions, and 4 additions of 99mTc-sodium pertechnetate were purposely added to create trials with RCP values below the accepted limit of 90% purity. **Results:** Two hundred four trials (100 of which were replicated) were run from the 17 kit preparations. Sixty-four (31%) of the 204 trials were below 90% purity based on the standard method. The overall agreement between the standard and alternative methods was 94% (192/204). The sensitivity of the alternative method for unacceptable RCP limits was 86% (55/64), and the specificity for acceptable RCP values was 98% (137/140). The agreement between the replicated trials of the alternative method was 99% (99/100), and for the standard method it was 92% (92/100). **Conclusion:** The standard method proved to be a much slower method and requires much more precision and attention. The alternative method is much faster, is easier, requires less attention to the solvent-development process, and can be used for RCP testing of both 99mTc-tetrofosmin and 99mTc-sestamibi. Furthermore, the stationary phase is much more readily available, is not moisture-sensitive, and is less susceptible to operator technique. Our method is accurate in determining the RCP value of 99mTc-tetrofosmin and is a better RCP testing method for 99mTc-tetrofosmin. * 99mTc-tetrofosmin * radiochemical purity * quality control * nuclear pharmacy * radiopharmaceutical The method suggested in the package insert for 99mTc-tetrofosmin (Myoview; GE Healthcare) to test the radiochemical purity (RCP) of the radiopharmaceutical uses instant thin-layer chromatography (ITLC) impregnated with silica gel paper (ITLC-SG) (Pall Corp.) as the stationary phase and a 35:65 acetone:dichloromethane mixture as the mobile phase (Fig. 1) (*1,2*). Because the ITLC-SG paper will no longer be commercially available, along with several other disadvantages of this method (Table 1), an alternative method is necessary. An alternative method for the RCP testing of 99mTc-tetrofosmin using Whatman 1 paper (*3*) currently exists, but we found that it took considerably longer than the 3 min indicated in the article (*3*). Additionally, after we evaluated the method, we found that it had a high specificity of 95% but a sensitivity of only 70% relative to the standard method (Eggert et al., unpublished data, 2009). ![FIGURE 1. ](http://tech.snmjournals.org/https://tech.snmjournals.org/content/jnmt/38/2/81/F1.medium.gif) [FIGURE 1. ](http://tech.snmjournals.org/content/38/2/81/F1) **FIGURE 1.**  Chromatography diagrams of standard (A) and alternative (B) methods for RCP testing of 99mTc-tetrofosmin. View this table: [TABLE 1](http://tech.snmjournals.org/content/38/2/81/T1) **TABLE 1** Comparison Between Standard and Alternative RCP Testing Systems An alternative RCP testing method for 99mTc-sestamibi, proposed by Hung et al. (*4*) in 1991, was evaluated for 99mTc-tetrofosmin to determine whether this same method could be used for both reconstituted products. The proposed alternative method uses solvent saturation pads (Krackeler Scientific, Inc.) as the stationary phase and a 1:1 chloroform:tetrahydrofuran mixture as the mobile phase (Fig. 1). The purpose of this research was to determine whether the alternative method was a reliable and valid method for testing the RCP of 99mTc-tetrofosmin. ## MATERIALS AND METHODS ### Standard RCP Testing Method for 99mTc-Tetrofosmin The manufacturer's suggestions call for ITLC-SG chromatography paper, which needs to be stored in a desiccator (*1*). The ITLC-SG paper must be cut into strips measuring 2 × 20 cm that are used as the stationary phase (*1*). Additionally, a 35:65 acetone:dichloromethane mixture is used as the mobile phase (Fig.1). The mixture should be poured into an ascending chromatography cylinder, which should be covered and allowed to equilibrate with the solvent vapor (*1*). The ITLC-SG strip must be marked accordingly to ensure accurate migration levels are achieved (Fig. 1). A 10- to 20-μL sample should be applied at the origin line using a 1-mL syringe with a 28.5-gauge needle. The strip should then be placed in the chromatography cylinder and covered immediately. To avoid any inaccuracies in the results, the strip should not be allowed to adhere to the walls of the cylinder. Once the solvent reaches the appropriate migration level, the strip should be removed from the tank and allowed to dry. The strip needs to then be cut into 3 pieces at the marked cut points. Each section should be measured separately using a dose calibrator to measure the activity present. In this system, free 99mTc-sodium pertechnetate runs to the top portion of the strip, 99mTc-tetrofosmin migrates to the center portion of the strip, and hydrolyzed-reduced 99mTc, along with any hydrophilic complex impurities, remains at the origin in the bottom portion of the strip. ### Alternative RCP Testing Method for 99mTc-Tetrofosmin The alternative method calls for solvent saturation pads, measuring 1×8.5 cm, to be used as the stationary phase (Fig.1). The strip should be marked accordingly to ensure proper migration levels are achieved (Fig. 1). A 10- to 20-μL sample must be placed at the origin using a 1-mL syringe with a 28.5-gauge needle. The sample strip should be placed into a blood collection tube (Kendall Monoject; Tyco Healthcare) with a 1:1 chloroform:tetrahydrofuran mixture acting as the mobile phase. The tube should be covered immediately after the introduction of the strip, and the strip should not be allowed to adhere to the walls of the tube. Once the solvent reaches the appropriate line, it should be removed from the chromatography tank and allowed to dry. On completion, the strip must be cut at the designated cut line, and the activity on each half of the strip should be measured using a dose calibrator. With this system, free 99mTc-sodium pertechnetate, hydrolyzed-reduced 99mTc, and any hydrophilic complex impurities remain at the origin in the bottom portion of the strip, and 99mTc-tetrofosmin runs to the top portion of the strip. ### Fabrication of Various RCP Levels of 99mTc-Tetrofosmin In determining whether the proposed alternative method was an accurate procedure, it was necessary to verify and compare the results of both the current standard ITLC-SG method and the proposed alternative method. Initially, each kit was prepared according to the package insert instructions (*1*). Using a shielded, sterile syringe, we added 8,880 MBq (240 mCi) of 99mTc generator eluate diluted to 8 mL into a shielded kit vial containing a venting needle. Before the syringe was removed from the vial, 2 mL of gas was withdrawn. The venting needle and syringe were then removed, and the vial was mixed gently for 10 s before being allowed to incubate at room temperature for 30 min, rather than 15 min as recommended in the package insert (*1*). The room temperature in our laboratory is set at 20°C rather than 25°C, because of the requirements of the United States Pharmacopeia (USP), General Chapter <797> (*5*). The colder temperature may slow down the transchelation process (*6*) during the reconstitution of 99mTc-tetrofosmin, which was the reason that the 30-min incubation period was used. After the initial RCP testing after 30 min of incubation, 99mTc-sodium pertechnetate was added to create preparations for trials with RCP values below 90% purity. A range of 296–444 MBq (8–12 mCi) of 99mTc-sodium pertechnetate was added to the reconstituted kit for each trial, and 4 additions—each within this specified range of radioactivity—were completed for each kit. After each addition, the kit was mixed gently and allowed to incubate at room temperature for 10 min before RCP testing was completed (Fig. 2). ![FIGURE 2. ](http://tech.snmjournals.org/https://tech.snmjournals.org/content/jnmt/38/2/81/F2.medium.gif) [FIGURE 2. ](http://tech.snmjournals.org/content/38/2/81/F2) **FIGURE 2.**  Two strips (initial strip and replicate) were run for both standard and alternative RCP testing methods at 30 min after initial reconstitution of kit at 0 min. Additions of 99mTc-sodium pertechnetate were made at 30, 40, 50, and 60 min. After each addition, there was 10 min of incubation, followed by RCP testing using both methods. RCP evaluation was completed at 40, 50, 60, and 70 min, each being a time point for separate 99mTc-sodium pertechnetate addition. Seventeen kits were prepared, and 204 trials were run. Of those trials, 100 were replicates. For each trial, 2 strips for each RCP testing method were completed. The second strip for each method was considered the replicate of that trial. The data were summarized, and the sensitivity and specificity of each method were determined. ### Comparison Between Standard and Alternative Testing Methods The RCP for each method was determined both before and after the fabrication of the kits. In doing this, 2 strips for each method were used to measure the RCP value of the prepared kit. ## RESULTS ### Fabrication of Various RCP Levels of 99mTc-Tetrofosmin The RCP of each reconstituted 99mTc-tetrofosmin vial (*n* = 17) was evaluated before and after the additions of 99mTc-sodium pertechnetate. 99mTc-sodium pertechnetate was added to generate a range of RCP values above and below the acceptable limits—values that were grouped as 60−89% (*n* = 64), 90−94% (*n* = 61), and 95−100% (*n* = 79). For evaluation purposes, a passing value for the radiochemical purity was greater than 90% and a failing value less than 90% (*1*). ### Comparison Between Standard and Alternative Testing Methods From the 17 kit preparations, 204 trials were conducted. Of those, 100 trials were replicates of the same kit preparation at each fabrication level. Of the total 204 trials, 64 (31%) were below the 90% purity based on the standard method. For the alternative method, the sensitivity for unacceptable RCP values (below 90% purity) was 86% (55/64). Additionally, the specificity for acceptable RCP values (greater than 90% purity) was 98% (137/140). The overall agreement between the standard and the alternative methods was 94% (192/204). The agreement between the replicated trials of the alternative method was 99% (99/100), and for the standard method it was 92% (92/100). ## DISCUSSION 99mTc-tetrofosmin is an important radiopharmaceutical in the assessment of regions of reversible myocardial ischemia and the evaluation of ventricular function (*7,8*). Because of the incubation period during the preparation of the 99mTc-tetrofosmin, a rapid RCP method is necessary to ensure the most prompt availability possible. The lengthy RCP method currently used, taking 20–30 min, is unacceptable in the fast-paced environment of nuclear pharmacy and nuclear medicine. The other more significant negative aspect of the standard RCP testing method is the availability of the stationary phase. Because the ITLC-SG paper is being discontinued and will no longer be commercially available, it is necessary that an alternative method be developed and analyzed for the future RCP testing of 99mTc-tetrofosmin. Simply the fact that the current method is approved by the Food and Drug Administration and USP does not mean that another method is not equally good or better (*9*). Although the standard method is USP-approved, USP also states that an alternative method may be used as long as it is evaluated and determined to be an accurate alternative method (*10*). The alternative RCP testing method is an expedited (2.3 ± 0.1 min) (*4*), user-friendly process that is both accurate and valid when compared with the standard ITLC method. Because the method is not as dependent on the operator's skill, obtaining accurate results is relatively simple. Another important factor to be considered in judging both RCP testing methods is the migration sensitivity between the 2 methods. With the standard method, it is imperative that the solvent front does not exceed the premarked line on the ITLC-SG paper. Otherwise, the RCP results could be altered, producing either false-positives or false-negatives. On the other hand, the migration of the solvent front with the alternative method is not as sensitive and is not affected if the solvent front exceeds the predefined level on the strip. Therefore, the alternative method will produce accurate results more often than the current method. When considering the 86% overall specificity, one could argue that the value is too low and a cause for concern, but it is important to look at the individual specificities for each method. Although the alternative method has a specificity of 99%, the standard method has a specificity of 92%, indicating that the standard method has lower test–retest reliability, and therefore the overall specificity of 86% is most likely due to the imprecision of the standard method. The alternative method is a more reliable process in the fast-paced environment of nuclear medicine, which is dependent on high-quality radiopharmaceuticals and accurate RCP testing. One concern with the alternative method involves the safety of the solvents used. There has been some concern as to whether the combination of chloroform and tetrahydrofuran might be harmful to the individual working with them (*11*). A possible solution to this problem could be combining the 2 ingredients in a fume hood to reduce any potential danger or threat that the solution could pose. Additionally, the mobile phase requires refrigeration and biweekly preparation, which can be a burdensome process. ## CONCLUSION Both the standard and alternative RCP methods have advantages and disadvantages. Yet the pros and cons of the alternative method outweigh those of the standard method, suggesting that the alternative method would be a better method for the RCP testing of 99mTc-tetrofosmin. The most important advantage of the alternative method over the standard method is the availability of the stationary phase. Also, the alternative method was more accurate in the determination of the RCP value of 99mTc-tetrofosmin and can be used to assess the RCP value of both 99mTc-tetrofosmin and 99mTc-sestamibi. ## Acknowledgments We thank Rose M. Busta for her assistance in the submission of this paper. We also thank Thomas J. Barna, CNMT; Robert A. Myhrvold, CNMT; and Megan Piatz, RT (R), for their assistance in our research at the Mayo Clinic Nuclear Pharmacy Laboratory. This paper was presented at the 56th Annual Meeting of the Society of Nuclear Medicine, Toronto, Canada, on June 15, 2009. Laura A. Eggert was selected by the SNM Technologist Section Awards Committee to receive the student award, radiopharmacy category, for her oral presentation. ## Footnotes * COPYRIGHT © 2010 by the Society of Nuclear Medicine, Inc. ## References 1. 1. Myoview [package insert]. Arlington Heights, IL: GE Healthcare; February 2006. 2. 2. Technetium Tc-99m tetrofosmin injection. In: *USP 32-NF27: United States Pharmacopeia and National Formulary.* Rockville, MD: United States Pharmacopeial Convention; 2009:3673–3674. 3. 3. McKay BF, Zimmer AM, Spies SM. Rapid miniaturized chromatography for technetium-99m-tetrofosmin. J Nucl Med Technol. 1997;25:52–54. [Abstract/FREE Full Text](http://tech.snmjournals.org/lookup/ijlink/YTozOntzOjQ6InBhdGgiO3M6MTQ6Ii9sb29rdXAvaWpsaW5rIjtzOjU6InF1ZXJ5IjthOjQ6e3M6ODoibGlua1R5cGUiO3M6NDoiQUJTVCI7czoxMToiam91cm5hbENvZGUiO3M6NDoiam5tdCI7czo1OiJyZXNpZCI7czo3OiIyNS8xLzUyIjtzOjQ6ImF0b20iO3M6MTg6Ii9qbm10LzM4LzIvODEuYXRvbSI7fXM6ODoiZnJhZ21lbnQiO3M6MDoiIjt9) 4. 4. Hung JC, Wilson ME, Brown ML, Gibbons RJ. Rapid preparation and quality control method for technetium-99m-2-methoxy isobutyl isonitrile (technetium-99m-sestamibi). J Nucl Med. 1991;32:2162–2168. [Abstract/FREE Full Text](http://tech.snmjournals.org/lookup/ijlink/YTozOntzOjQ6InBhdGgiO3M6MTQ6Ii9sb29rdXAvaWpsaW5rIjtzOjU6InF1ZXJ5IjthOjQ6e3M6ODoibGlua1R5cGUiO3M6NDoiQUJTVCI7czoxMToiam91cm5hbENvZGUiO3M6Njoiam51bWVkIjtzOjU6InJlc2lkIjtzOjEwOiIzMi8xMS8yMTYyIjtzOjQ6ImF0b20iO3M6MTg6Ii9qbm10LzM4LzIvODEuYXRvbSI7fXM6ODoiZnJhZ21lbnQiO3M6MDoiIjt9) 5. 5. Pharmaceutical compounding-sterile preparations (general chapter 797). In: *USP 32-NF27: United States Pharmacopeia and National Formulary.* Rockville, MD: United States Pharmacopeial Convention; 2009:319–336. 6. 6. Kowalsky RJ, Falon SW. Radiopharmaceutical chemistry. In: *Radiopharmaceuticals in Nuclear Pharmacy and Nuclear Medicine*, 2nd ed. Washington, DC: American Pharmacists Association; 2004:235–336. 7. 7. Pandit-Taskar N, Grewal RK, Strauss HW. Cardiovascular system. In: Christian PE, Waterstram-Rich KM, eds. *Nuclear Medicine and PET/CT Technology and Techniques.* 6th ed. St. Louis, MO: Mosby; 2007:479–512. 8. 8. Saha GB. Characteristics of specific radiopharmaceuticals. In: *Fundamentals of Nuclear Pharmacy.* 5th ed. New York, NY: Springer; 2004:111–150. 9. 9. Hung JC, Ponto JA, Gadient KR, et al. Deficiencies of product labeling directions for the preparation of radiopharmaceuticals. J Am Pharm Assoc. 2004;44:30–35. [CrossRef](http://tech.snmjournals.org/lookup/external-ref?access_num=10.1331/154434504322713200&link_type=DOI) 10. 10. General notices and requirements. In: *USP 32-NF27: United States Pharmacopeia and National Formulary.* Rockville, MD: United States Pharmacopeial Convention; 2009:3–12. 11. 11. Luebke AL, Wilary DM, Mahoney DW, Hung JC. Evaluation of an alternative radiochemical purity testing method for technetium-99m-sestamibi. J Nucl Med Technol. 2000;28:259–263. [Abstract/FREE Full Text](http://tech.snmjournals.org/lookup/ijlink/YTozOntzOjQ6InBhdGgiO3M6MTQ6Ii9sb29rdXAvaWpsaW5rIjtzOjU6InF1ZXJ5IjthOjQ6e3M6ODoibGlua1R5cGUiO3M6NDoiQUJTVCI7czoxMToiam91cm5hbENvZGUiO3M6NDoiam5tdCI7czo1OiJyZXNpZCI7czo4OiIyOC80LzI1OSI7czo0OiJhdG9tIjtzOjE4OiIvam5tdC8zOC8yLzgxLmF0b20iO31zOjg6ImZyYWdtZW50IjtzOjA6IiI7fQ==) * Received for publication October 22, 2009. * Accepted for publication February 4, 2010.