JNMT
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Folks, R. D.
Right arrow Articles by Taylor, A. T.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Folks, R. D.
Right arrow Articles by Taylor, A. T.
Journal of Nuclear Medicine Technology Volume 35, Number 1, 2007 27-33
© 2007 by Society of Nuclear Medicine

Development and Prospective Evaluation of an Automated Software System for Quality Control of Quantitative 99mTc-MAG3 Renal Studies

Russell D. Folks, CNMT, RT (N), Ernest V. Garcia, PhD and Andrew T. Taylor, MD

Emory University School of Medicine, Atlanta, Georgia

Correspondence: For correspondence or reprints contact: Russell D. Folks, CNMT, RT (N), Division of Nuclear Medicine, Emory University School of Medicine, 1364 Clifton Rd. NE, Atlanta, GA 30322. E-mail: rfolks{at}emory.edu

Quantitative nuclear renography has numerous potential sources of error. We previously reported the initial development of a computer software module for comprehensively addressing the issue of quality control (QC) in the analysis of radionuclide renal images. The objective of this study was to prospectively test the QC software. Methods: The QC software works in conjunction with standard quantitative renal image analysis using a renal quantification program. The software saves a text file that summarizes QC findings as possible errors in user-entered values, calculated values that may be unreliable because of the patient's clinical condition, and problems relating to acquisition or processing. To test the QC software, a technologist not involved in software development processed 83 consecutive nontransplant clinical studies. The QC findings of the software were then tabulated. QC events were defined as technical (study descriptors that were out of range or were entered and then changed, unusually sized or positioned regions of interest, or missing frames in the dynamic image set) or clinical (calculated functional values judged to be erroneous or unreliable). Results: Technical QC events were identified in 36 (43%) of 83 studies. Clinical QC events were identified in 37 (45%) of 83 studies. Specific QC events included starting the camera after the bolus had reached the kidney, dose infiltration, oversubtraction of background activity, and missing frames in the dynamic image set. Conclusion: QC software has been developed to automatically verify user input, monitor calculation of renal functional parameters, summarize QC findings, and flag potentially unreliable values for the nuclear medicine physician. Incorporation of automated QC features into commercial or local renal software can reduce errors and improve technologist performance and should improve the efficiency and accuracy of image interpretation.

Key Words: renal function; quality control; 99mTc-MAG3; MAG3 clearance




This article has been cited by other articles:


Home page
JNMHome page
A. Taylor, E. V. Garcia, J. N. G. Binongo, A. Manatunga, R. Halkar, R. D. Folks, and E. Dubovsky
Diagnostic Performance of an Expert System for Interpretation of 99mTc MAG3 Scans in Suspected Renal Obstruction
J. Nucl. Med., February 1, 2008; 49(2): 216 - 224.