JNMT
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Kowalsky, R. J.
Right arrow Articles by Johnston, R. E.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Kowalsky, R. J.
Right arrow Articles by Johnston, R. E.

Journal of Nuclear Medicine Technology, Vol 26, Issue 2 94-98, Copyright © 1998 by Society of Nuclear Medicine


ARTICLES

Dose calibrator assay of iodine-123 and indium-111 with a copper filter

RJ Kowalsky and RE Johnston
University of North Carolina, Department of Radiology, Chapel Hill 27599, USA.

OBJECTIVE: The accurate measurement of radionuclides in a dose calibrator can be affected by container geometry especially for low-energy photons. This study investigated the effectiveness of using copper filtration for selective absorption of low-energy x-rays during the measurement of 123I and 111In to minimize variations in measurement as a function of geometry. METHODS: Copper absorbers of varying thickness were used to define the absorption profile of low-energy x-rays in the 20- to 50-keV energy window with 125I and a gamma scintillation spectrometer. Iodine-123 and 111In standards were assayed in a dose calibrator, with and without a copper filter positioned in the ion chamber well, to determine new isotope calibration settings for these nuclides with the copper filter. Samples of 123I and 111In, of constant activity but various volumes, were assayed in 1-ml, 3-ml and 10-ml plastic syringes and a 10-ml glass vial with and without copper filtration to assess the influence of geometry on the accuracy of measurement. RESULTS: A copper thickness of 0.6 mm removed 99.5% of 125I photons in the 20- to 50-keV window and significantly reduced the low-energy component of 123I and 111In. Activity measurements of 123I and 111In in the dose calibrator varied with container type and volume without copper filtration but were essentially unchanged when a copper filter was used. CONCLUSION: A copper filter > or = 0.6 mm thick eliminates the need for geometry correction factors for measuring 123I and 111In with the dose calibrator.


This article has been cited by other articles:


Home page
J. Nucl. Med. Technol.Home page
J.-F. Valley, S. Bulling, M. Leresche, and C. Wastiel
Determination of the Efficiency of Commercially Available Dose Calibrators for {beta}-Emitters
J. Nucl. Med. Technol., March 1, 2003; 31(1): 27 - 32.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
THE JOURNAL OF NUCLEAR MEDICINE JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY
Copyright © 1998 by the Society of Nuclear Medicine Technologist Section.